Accuracy and precision of curcumol in rat plasma and tissue homogenates (n = 5).

Sample Spiked concentration (ng/mL) Intra-day Inter-day
Accuracy Precision Accuracy Precision
% (RSD, %) % (RSD, %)
Plasma 10 105.9 8.2 100.6 7.5
30 102.2 8.4 96.5 8.2
400 110.9 2.0 98.3 9.8
1,600 111.0 2.3 97.3 10.7
Brain 5 100.0 12.3 94.0 8.8
15 98.0 2.1 95.7 4.5
160 97.3 1.7 98.3 1.7
1,600 93.6 4.4 92.5 2.9
Heart 5 92.4 2.5 93.8 2.8
15 90.9 3.4 93.2 3.1
160 88.0 2.2 99.3 8.5
1,600 97.1 4.5 94.0 4.1
Liver 5 96.4 5.9 102.4 6.9
15 99.1 8.6 104.4 6.9
160 98.6 3.4 106.3 5.9
1,600 100.3 3.1 98.4 5.6
Spleen 5 94.6 6.7 97.6 8.1
15 104.1 7.5 102.4 6.6
160 108.5 1.1 100.4 6.3
1,600 101.5 5.6 96.6 6.1
Lung 5 96.4 7.6 97.3 7.5
15 102 4.4 96.4 5.8
160 107.1 1.7 99.3 8.4
1,600 89.1 0.9 93.4 7.3
Kidney 5 94.9 7.3 97.3 8.0
15 95.1 6.7 99.0 6.5
160 94.9 4.7 101.2 8.4
1,600 89.8 2.9 96.3 9.6
Intestine 10 100.6 8.6 102.0 7.4
30 100.7 6.0 102.9 4.7
400 111.0 2.5 108.4 4.9
4,000 94.1 3.4 95.9 3.9
Colon 10 109.4 3.5 102.1 8.0
30 103.1 6.6 105.0 5.3
400 99.0 1.2 106.5 5.5
4,000 90.9 3.6 96.6 5.9