Table 4. Statistical Comparison of the Recommended Method with UV–Vis Reference Method for Fluoride Determination.

method mean conc. (mg L–1) SD (σ) Sa,b ta,b ttablob Fb Ftablob
recommended method 12.621 0.134          
UV–Vis reference method 12.551 0.151 0.169 0.291 2.306 1.591 6.39
a

S2 = ((n1 – 1)s12 + (n2 – 1)s22)/(n1 + n2 – 2) and t = (12)/(S(1/n1 + 1/n2)1/2), where S is the pooled standard deviation, s1 and s2 are the standard deviations of the two populations with sample sizes of n1 and n2, and 1 and 2 are sample means (t has (n1 + n2 – 2) degrees of freedom); here, n1 = n2 = 5.

b

Statistical comparison made on paired data produced with the proposed and reference methods; results given only on the row of the reference method.