Confusion matrices when a 1% machine learning model threshold is used to predict a 3% IC threshold

All machines
n = 603 Ion chamber Plans to QA
Fail Pass
Machine Learning model Fail 41 257 49.4%
Pass 0 305
Truebeam STx
n = 141 Ion chamber
Fail Pass
Machine Learning model Fail 6 31 26.2%
Pass 0 104
Truebeam
n = 462 Ion chamber
Fail Pass
Machine Learning model Fail 35 226 56.5%
Pass 0 201

Abbreviations: IC = ion chamber; QA = quality assurance.