Table 1.
Performance Metrics of the Algorithm With Various Modifications, Including Adjustment of the Delta Factor and the Addition of Upper and Lower Limits
| Delta | No Upper and Lower Thresholds | Lower (> 30 mm Hg) and Upper (< 200 mm Hg) Thresholds | ||||
|---|---|---|---|---|---|---|
| Sensitivity (%) | Specificity (%) | MCC | Sensitivity (%) | Specificity (%) | MCC | |
| 1× deltas | 92.31 | 98.75 | 0.39 | 97.48 | 98.59 | 0.40 |
| 0.4× deltas | 94.96a | 98.02a | 0.34 | 97.48 | 98.02 | 0.35 |
| 1.3× deltas | 93.28b | 98.68b | 0.40 | 97.48a | 98.68a | 0.41 |
| 2× deltas | 91.60 | 98.72 | 0.40 | 96.64b | 98.72b | 0.42 |
| 4× deltas | 86.55 | 98.72 | 0.37 | 95.80 | 98.72 | 0.41 |
MCC = Matthew correlation coefficient.
aSensitivity and specificity optimized.
bMCC optimized
.