Table 5.
Diagnostic performance of FFR, FFR, and FFR to detect ischemic stenoses at the clinical threshold of 0.80.
| Metric | FFR | FFR | FFR |
|---|---|---|---|
| Sensitivity | 79.0 (54.4–94.0) | 89.5 (66.9–98.7) | 89.5 (66.9–98.7) |
| Specificity | 90.3 (74.3–98.0) | 93.6 (78.6–99.2) | 93.6 (78.6–99.2) |
| PPV | 83.3 (62.5–93.8) | 89.5 (68.8–97.0) | 89.5 (68.8–97.0) |
| NPV | 87.5 (74.4–94.4) | 93.6 (79.6–98.2) | 93.6 (79.6–98.2) |
| Overall accuracy | 86.0 (73.3–94.2) | 92.0 (80.8–97.8) | 92.0 (80.8–97.8) |