Skip to main content
. 2022 Dec 20;290:122265. doi: 10.1016/j.saa.2022.122265

Table 1.

Regression and validation data for the determination of RDV and APX by the proposed method.

Parameters RDV APX
Wavelength (nm) 410 469
Linearity range (ng/mL) 5–200 50–3000
LOD (ng/mL) 0.603 8.598
LOQ (ng/mL) 1.828 26.054
Slope 0.0349 0.0023
Intercept 0.2604 0.3678
Coefficient of determination (r2) 0.9992 0.9994
Accuracy (%Er)a −0.14 0.59
Repeatability Precision (RSD)b 0.771 0.624
Intermediate precision (RSD)b 0.925 0.815
a

The mean average percentage error of nine determinations (three concentrations repeated three times).

b

RSD of nine determinations (three concentrations repeated three times).