TABLE 2.
IgM antibody detection by DPP HIV-Syphilis assay in active syphilisa
| Sample ID | Days since first bleed | RPR |
Testing prior to 2-ME treatment |
DPP signal reduction after 2-ME treatment |
|||||
|---|---|---|---|---|---|---|---|---|---|
| DPP HIV-Syphilis |
DPP HIV-Syphilis IgM |
DPP HIV-Syphilis | DPP HIV-Syphilis IgM | ||||||
| Result | Titer | Resultb | s/coc | Resultb | s/coc | ||||
| PSS901-06 | 45 | Positive | 1:8 | Positive | 3.7 | Positive | 1.4 | 95% | 99% |
| PSS901-07 | 48 | Positive | 1:16 | Positive | 4.7 | Positive | 2.0 | 87% | 99% |
| PSS901-08 | 52 | Positive | 1:32 | Positive | 16.0 | Positive | 2.6 | 84% | 98% |
| PSS901-09 | 59 | Positive | 1:64 | Positive | 33.9 | Positive | 3.6 | 58% | 95% |
| w36981501276400 | NA | Positive | 1:1 | Positive | 5.4 | Positive | 1.4 | 91% | 100% |
| w36981501452200 | NA | Positive | 1:16 | Positive | 30.7 | Positive | 2.6 | 68% | 97% |
| w36981502181900 | NA | Positive | 1:1 | Positive | 30.9 | Positive | 1.4 | 62% | 99% |
| PRB-933-02 | 21 | ND | NA | Positive | 1.3 | Positive | 1.2 | 87% | 98% |
| DLS0116358 | NA | Positive | 1:64 | Positive | 36.6 | Negative | 0.9 | 16% | NA |
| PRB-912-02 | 9 | ND | NA | Positive | 5.7 | Negative | 0.3 | 8% | NA |
RPR, rapid plasma reagin; s/co, signal-to-cutoff ratio; ND, not done; NA, not applicable.
Results are shown for treponemal antibodies, all samples were HIV antibody negative, except for PRB-933-02 and PRB-912-02.
Test interpretation: positive result at signal-to-cutoff ration (s/co) ≥1.0, negative result at s/co <1.0.