. 2022 Oct 31;10(6):e03125-22. doi: 10.1128/spectrum.03125-22

TABLE 4.

Target viral load of specimens

Viral load (log10 IU or copies per mL) (N per site)
Study HBV HCV HIV-1
Method comparison Negative (30) Negative (30) Negative (30)
1.0–3.7 (50)a 1.2–3.5 (50) 1.3–3.3 (50)
3.7–6.3 (50) 3.5–5.7 (50) 3.3–5.0 (50)
6.3–9.0 (50) 5.7–8.0 (50) 5.0–7.0 (50)
Reproducibility 1.0 (LLOQ) (30) 1.2 (LLOQ) (30) 1.3 (LLOQ) (30)
1.7 (30) 1.4 (30) 1.7 (30)c
3.0 (30) 2.0 (30) 2.3 (30)
3.3 (30) 4.0 (30) 3.0 (30)
4.3 (30) 5.9 (30) 5.0 (30)
5.3 (30) 6.8 (30) 5.7 (30)
9.0 (ULOQ) (30)b,d 8.0 (ULOQ) (30) 7.0 (ULOQ) (30)
a

Number of specimens for method comparison in each VL range counted based on cobas 6800 result at site 3.

b

Nine pairs of results were excluded from analysis because one result of each pair was outside the linear range (over the ULOQ).

c

One pair of results was excluded from analysis because one result was outside the linear range (below the ULOQ).

d

LLOQ, lower limit of quantitation; ULOQ, upper limit of quantitation.