Table 4.

Precision results of double examinations of the femoral and tibial components

Factor Femoral component (n = 11) Tibial component (n = 17)
Mean (SD) PI Mean (SD) PI
Translation, mm
 X-axis 0.013 (0.12) 0.23 –0.019 (0.074) 0.15
 Y-axis –0.012 (0.038) 0.074 –0.010 (0.057) 0.11
 Z-axis –0.002 (0.22) 0.43 0.006 (0.10) 0.31
Rotation, degree
 X-axis 0.046 (0.097) 0.19 0.072 (0.40) 0.79
 Y-axis –0.036 (0.32) 0.62 –0.035 (0.25) 0.48
 Z-axis –0.060 (0.15) 0.30 0.094 (0.29) 0.57
MTPM, mm 0.26 (0.26) 0.28 (0.20)

PI = precision interval (1.96*SD). In case of non-normality only mean (SD) is presented.