Table 3.
First recording (N = 129) | Second recording (N = 129) | Third recording (N = 129) | First recording incl. unclassifiable results (N = 129) | First recording excl. patients with CIED (N = 104) | First recording excl. patients with paced rhythm (N = 113) | |
---|---|---|---|---|---|---|
Sensitivity, AF | 93.5% | 94.6% | 93.0 % | 66.2% | 97.5% | 95.5% |
Specificity, SR | 100% | 100% | 96.5 % | 73.4% | 100% | 100% |
% Inconclusive | 6.2% | 3.9% | 3.9% | - | 5.8% | 5.3% |
% Poor recording | 18.6% | 7.8% | 4.7% | - | 17.3% | 17.7% |
AF = atrial fibrillation; CIED = cardiac implantable electronic device; Excl. = excluding; HR = heart rate; Incl. = including; SR = sinus rhythm.