Overview of the influence of repeated measurements, including unclassifiable results, implanted cardiac device, and paced rhythms, on the accuracy of the single-lead electrocardiographic Apple Watch notification

First recording (N = 129) Second recording (N = 129) Third recording (N = 129) First recording incl. unclassifiable results (N = 129) First recording excl. patients with CIED (N = 104) First recording excl. patients with paced rhythm (N = 113)
Sensitivity, AF 93.5% 94.6% 93.0 % 66.2% 97.5% 95.5%
Specificity, SR 100% 100% 96.5 % 73.4% 100% 100%
% Inconclusive 6.2% 3.9% 3.9% - 5.8% 5.3%
% Poor recording 18.6% 7.8% 4.7% - 17.3% 17.7%

AF = atrial fibrillation; CIED = cardiac implantable electronic device; Excl. = excluding; HR = heart rate; Incl. = including; SR = sinus rhythm.