Table 1.

Summary of point-of-care thermal ablation devices

          Liger           MedGyn           Wisap
Model No.
          HTU-110
          MTA-100
          C3
Cost
          US$1500 (includes 4 probes)*
          US$1550 (includes one tip)*
          €2326 ( ~ US$2900) (includes 2 probes and 2 sliders)*
Cost of additional probes/tips
          US$100 each*
          $245 USD each*
          €602 each ( ~ US$725)*
Power sources available
          Rechargeable battery
          Rechargeable battery
          Wall plug or rechargeable battery (power supply included)
Number of batteries/chargers included
          2 batteries, 1 charger
          2 batteries, 1 charger
          1 battery, 1 charger
Number of cycles per battery charge
           ~ 30-60
           ~ 100 (depends on length of treatment cycle)
           ~ 140
Low battery indicator
          Yes
          Yes
          Yes
Probe treatment temperature
          100°C
          100°C
          100°C
Probe heating time
          8 s
          30-50 s
          20-30 s
Treatment time
          20-60 s (beeps after 1/4, 1/2, and 3/4 of treatment time)†
          20, 40, or 60 s†
          60 s (beeps after 30, 45 and 60 s)
Built in timer
          Yes
          Yes
          Yes
Light source
          LED illumination
          LED illumination
          LED illumination
Temperature display
          No‡
          Yes
          No
Probe options
          16mm flat, 19mm flat, 19mm nipple
          19mm flat, 19mm x 5mm convex, 19mm x 10mm endocervical, 25mm x 7mm endocervical, 25mm flat
          17 mm flat, 20 mm flat, 20 mm nipple
Removable probes
          Yes
          Only probe tip is removable, shaft is not removable
          Yes
Heat protection
          Yes, via an integrated plastic ring that covers the heating element.
          Yes, via an integrated plastic ring that covers the heating element in the tip.
          Yes, via a retractable sleeve that covers the heating element.
Probe lifetime
           ~ 100 cycles
          at least 50 cycles
           ~ 500 cycles
Method of disinfection           Removable probe can be soaked in Cidex or autoclaved. Autoclaving may reduce lifetime of probe.           Hand unit and shaft can be cleaned with CaviWipes, Cavicide, and enzymatic detergent, followed by wiping with damp cloth. Care should be taken to avoid getting water inside the unit. The tips can be autoclaved.           Removable probe can be soaked in Cidex. Probes should not be autoclaved.

HTU-110 – Liger thermocoagulator, MTA-100 – Medgyn thermal ablation system, C3 – WISAP cervical cold coagulator

*Pricing should be verified with manufacturer; discounts may be available when buying in volume.

†Treatment time is programmable.

‡Temperature is internally monitored; error signal will be generated if temperature is out of range.