Table 3.
Manufacturer and Model | No. of Patients | No. of Scans | Correlationa | Total Kidney Volume Difference,b ml | Total Kidney Volume Difference, % | P Valuec |
---|---|---|---|---|---|---|
Siemens | ||||||
Aera | 43 | 109 | 0.991 | −67±185 | −1.9±9.9 | 0.16 |
Avanto | 41 | 113 | 0.992 | −47±160 | −1.9±7.8 | 0.07 |
SymphonyTim | 29 | 72 | 0.991 | −52±200 | −1.3±8.7 | 0.58 |
Avanto fit | 12 | 23 | 0.995 | −29±306 | 1.3±5.8 | 0.37 |
Amira | 8 | 25 | 0.990 | 2±81 | 0.5±5.0 | 0.84 |
Other models | 43 | 92 | 0.994 | −63±167 | −1.9±7.1 | 0.01 |
Philips | ||||||
Ingenia | 26 | 68 | 0.998 | −41±105 | 1.4±5.8 | 0.37 |
Intera | 31 | 80 | 0.995 | −116±550 | −2.3±6.3 | 0.07 |
Achieva | 24 | 66 | >0.999 | −56±134 | −1.1±3.9 | 0.05 |
Ingenia S | 14 | 40 | 0.999 | −39±75 | 2.0±4.6 | 0.17 |
Achieva dStream | 7 | 20 | >0.999 | −10±27 | 0.3±3.1 | 0.69 |
Panorama HFO | 8 | 20 | 0.997 | −52±281 | 2.5±11.1 | >0.99 |
GE | ||||||
Multiple models | 29 | 73 | 0.997 | −31±108 | −0.9±5.8 | 0.001 |
TOSHIBA | ||||||
Multiple models | 5 | 12 | 0.999 | 149±378 | −0.1±11.0 | 0.81 |
High correlations and consistently small differences from reference volumetry can be noticed for the automated method across all of the models.
Pearson correlation coefficient.
Total kidney volume difference was calculated between automated and manual methods.
Calculated with the Wilcoxon signed-rank test.