TABLE 4.

Amplitude values of axial and sagittal measurements, ICCs, and Bland‐Altman coefficients in each cervical segment in subgroups with different venc

Velocity encoding 2 cm/second
Axial Sagittal Bland‐Altman coefficients Intraclass correlation coefficients
Segment N Mean SD Mean SD Mean SD 1.96 × SD p ICC CImin CImax p
C2 28 0.54 0.33 0.50 0.27 0.04 0.14 0.28 .362 .882 .762 .943 <.001
C3 10 0.97 0.64 0.88 0.62 0.10 0.36 0.70 .285 .843 .509 .958 .001
C4 19 1.05 0.59 0.96 0.58 0.09 0.26 0.52 .165 .894 .749 .958 <.001
C5 22 1.05 0.53 1.03 0.49 0.02 0.21 0.42 .638 .915 .808 .964 <.001
C6 10 1.00 0.25 0.96 0.24 0.04 0.19 0.36 .559 .732 .235 .926 .006
Velocity encoding 3 cm/second
Axial Sagittal Bland‐Altman coefficients Intraclass correlation coefficients
Segment N Mean SD Mean SD Mean SD 1.96 × SD p ICC CImin CImax p
C2 34 0.47 0.22 0.46 0.24 0.01 0.09 0.17 .358 .931 .867 .965 <.001
C3 33 0.79 0.53 0.72 0.53 0.07 0.11 0.21 .001 .973 .914 .989 <.001
C4 32 0.70 0.34 0.93 0.51 –0.24 0.25 0.48 <.001 .734 .167 .899 <.001
C5 32 1.05 0.55 1.04 0.56 0.00 0.17 0.33 .921 .956 .912 .978 <.001
C6 31 0.86 0.51 0.85 0.45 0.01 0.23 0.45 .767 .887 .780 .944 <.001

Note: The motion amplitude values for the subgroups with venc 2 cm/second (upper table) and venc 3 cm/second (lower table) measurements are reported per cervical segment for axial and sagittal measurements. The amplitude difference of axial minus sagittal measurements was significantly different from 0 at segment C3 and C4 in the venc 3 cm/second subgroup. Intraclass correlation coefficients (ICCs; two‐way mixed model, single measures, absolute agreement) with its 95% confidence interval (CI) are presented for each cervical segment.

Abbreviations: N, number of measurements; SD, standard deviation.