Table 3.

Accuracy of PeekNV as a NVI Screening Test

Parameter Point Estimate Lower Bound 95% CI Upper Bound 95% CI
Sensitivity 89.61% 85.19% 94.03%
Specificity 95.28% 92.21% 98.35%
Positive predictive value 93.24% 89.61% 96.88%
Negative predictive value 92.66% 88.88% 96.44%
Diagnostic accuracy 92.90% 88.23% 95.80%
Cohen's kappa (unweighted) 0.85 0.708 0.998

NVI defined as vision worse than N6 at 40 cm. The reference test was the conventional Tumbling ā€œEā€ chart. These results are for the right-eye-first examination only, as when used in other contexts the test will usually only be used once, without practice effect.