Table 3.
Accuracy of PeekNV as a NVI Screening Test
Parameter | Point Estimate | Lower Bound 95% CI | Upper Bound 95% CI |
---|---|---|---|
Sensitivity | 89.61% | 85.19% | 94.03% |
Specificity | 95.28% | 92.21% | 98.35% |
Positive predictive value | 93.24% | 89.61% | 96.88% |
Negative predictive value | 92.66% | 88.88% | 96.44% |
Diagnostic accuracy | 92.90% | 88.23% | 95.80% |
Cohen's kappa (unweighted) | 0.85 | 0.708 | 0.998 |
NVI defined as vision worse than N6 at 40 cm. The reference test was the conventional Tumbling āEā chart. These results are for the right-eye-first examination only, as when used in other contexts the test will usually only be used once, without practice effect.