. 2022 Dec 13;35(6):321–332. doi: 10.1089/jamp.2022.0005

Table 1.

Andersen Cascade Impactor Findings for INP104

Lot Device storage orientation Parameter Result
1 Upright % <9 μm 0.30%
Total recovery (%LC) 97.18%
Inverted % <9 μm 0.12%
Total recovery (%LC) 98.22%
2 Upright % <9 μm 0.08%
Total recovery (%LC) 99.06%
Inverted % <9 μm 0.10%
Total recovery (%LC) 102.86%

The majority of droplets in INP104 as detected by ACI are outside the respirable range (conservatively defined here as 9 microns), regardless of whether the device is stored in the upright or inverted orientation.

ACI, Andersen Cascade Impactor; LC, label claim.