Table 2.
RM 8012 | RM 8013 | |
---|---|---|
n | 62 | 66 |
reported diameter (ROI)c | 95.9 ± 5.7 (11.5)d | 106.0 ± 2.0 (4.0)d |
HR-SEM diameterc | 98.4 ± 19.3 (38.7)d | 102.7 ± 10.5 (21.0)d |
spICP-MS diameterc | 100.4 ± 4.7 (9.4)d | 97.7 ± 22.2 (44.4)d |
TEM (ROI) full PSD | 90.4 ± 5.5 | 102.1 ± 2.5 |
HR-SEM full PSD | 96.5 ± 18.5 (37.0)d | 100.8 ± 10.1 (20.1)d |
spICP-MS full PSD | 86.1 ± 4.7 (9.4)d | 93.9 ± 20.8 (41.5)d |
spICP-MS simultaneous diameterc | 96.9 ± 1.7 (3.3)d | 103.9 ± 1.5 (3.1)d |
spICP-MS simultaneous full PSD | 87.0 ± 2.3 (4.6)d | 94.4 ± 1.5 (3.0)d |
Results are expressed as a ratio of PNCdirect to PNCmean or PNCdistribution multiplied by 100, where PNCmean or PNCdistribution was derived based on the combination of Au mass fraction with the mean particle diameter or PSD reported by TEM,31,33 HR-SEM,32 spICP-MS,32 and the particle size results simultaneously obtained by spICP-MS in this study as specified in each column heading.
Values indicate the ratio between PNCdirect and various derived PNCs, expressed in percentage, and the combined standard uncertainty (uc) associated.
Assumes all analyte is present as spherical NPs of the central tendency diameter.
All recognized and evaluated sources of bias affecting PNC determinations were included in the parenthetical expanded uncertainty computation that corresponds to the expanded uncertainty for 95% coverage.