Published in final edited form as: Anal Chem. 2022 Feb 10;94(7):3091–3102. doi: 10.1021/acs.analchem.1c04140

Table 2.

Comparison between Average PNCdirect Obtained across 15 Independent spICP-MS Experiments for RM 8012 and RM 8013 and Various Derived PNCsa,b

RM 8012 RM 8013
n 62 66
reported diameter (ROI)c 95.9 ± 5.7 (11.5)d 106.0 ± 2.0 (4.0)d
HR-SEM diameterc 98.4 ± 19.3 (38.7)d 102.7 ± 10.5 (21.0)d
spICP-MS diameterc 100.4 ± 4.7 (9.4)d 97.7 ± 22.2 (44.4)d
TEM (ROI) full PSD 90.4 ± 5.5 102.1 ± 2.5
HR-SEM full PSD 96.5 ± 18.5 (37.0)d 100.8 ± 10.1 (20.1)d
spICP-MS full PSD 86.1 ± 4.7 (9.4)d 93.9 ± 20.8 (41.5)d
spICP-MS simultaneous diameterc 96.9 ± 1.7 (3.3)d 103.9 ± 1.5 (3.1)d
spICP-MS simultaneous full PSD 87.0 ± 2.3 (4.6)d 94.4 ± 1.5 (3.0)d
a

Results are expressed as a ratio of PNCdirect to PNCmean or PNCdistribution multiplied by 100, where PNCmean or PNCdistribution was derived based on the combination of Au mass fraction with the mean particle diameter or PSD reported by TEM,31,33 HR-SEM,32 spICP-MS,32 and the particle size results simultaneously obtained by spICP-MS in this study as specified in each column heading.

b

Values indicate the ratio between PNCdirect and various derived PNCs, expressed in percentage, and the combined standard uncertainty (uc) associated.

c

Assumes all analyte is present as spherical NPs of the central tendency diameter.

d

All recognized and evaluated sources of bias affecting PNC determinations were included in the parenthetical expanded uncertainty computation that corresponds to the expanded uncertainty for 95% coverage.