Table 3.

spICP-MS Determination of PNCdirect and Ratio between PNCdirect and PNCmean, Expressed in Percentage, for Different Commercial AuNPsa

measured PNCdirect (L−1)b supplier diameter + supplier Au mass fractionc (%) supplier diameter + in-house Au mass fractionc (%) HR-SEM diameter + in-house Au mass fractionc (%) spICP-MS diameter + in-house Au mass fractionc (%)
30 nm PVP (2.19 ± 0.20) × 1014 117.4 ± 7.6 95.0 ± 6.9 92.2 ± 16.7 (33.4)d 97.0 ± 6.6 (13.1)d
30 nm bPEI (2.03 ± 0.36 )X 1014 115.1 ± 12.8 96.4 ± 9.8 102.8 ± 18.5 (37.0)d 98.3 ± 8.8 (17.7)d
30 nm PEG (2.16 ± 0.30) × 1014 148.4 ± 13.6 131.6 ± 10.6 101.4 ± 17.0 (34.1)d 108.9 ± 7.4 (14.7)d
60 nm PVP (2.69 ± 0.24) × 1013 89.9 ± 9.1 81.2 ± 8.9 96.4 ± 11.1 (22.2)d 91.7 ± 6.3 (12.7)d
60 nm bPEI first lot (2.46 ± 0.28) × 1013 127.5 ± 12.2 111.3 ± 14.7 91.4 ± 13.1 (26.1)d 90.8 ± 10.3 (20.6)d
60 nm bPEI second lot (2.80 ± 0.37) × 1013 125.0 ± 12.3 103.9 ± 9.0 90.9 ± 9.7 (19.4)d 90.3 ± 5.6 (11.1)d
60 nm PEG first lot (2.99 ± 0.41) × 1013 160.0 ± 14.6 148.4 ± 11.1 101.8 ± 9.4 (18.7)d 96.3 ± 2.3 (4.5)d
60 nm PEG second lot (2.65 ± 0.45) × 1013 144.1 ± 13.9 126.8 ± 10.6 100.1 ± 10.4 (20.8)d 93.2 ± 8.2 (16.5)d
100 nm citrate (5.75 ± 0.73) × 1012 129.4 ± 13.8 130.6 ± 14.0 108.8 ± 11.2 (22.3)d 91.4 ± 7.6 (15.2)d
100 nm PVP (4.68 ± 0.74) × 1012 91.6 ± 7.8 107.4 ± 8.4 93.0 ± 8.9 (17.8)d 93.0 ± 8.0 (16.0)d
100 nm bPEI (4.93 ± 0.59) × 1012 93.4 ± 8.4 115.8 ± 12.9 104.7 ± 11.9 (23.8)d 99.9 ± 9.8 (19.7)d
100 nm PEG (6.33 ± 0.59) × 1012 139.2 ± 13.3 126.9 ± 10.3 90.9 ± 6.6 (13.3)d 104.3 ± 6.3 (12.7)d
a

PNCmean values reported for comparison to PNCdirect were derived based on the combination of Au mass fraction and the mean particle size reported by TEM,34 HR-SEM,32 and spICP-MS.32 With regard to Au mass fraction, values provided by the supplier were used in the first column while in-house determinations were used for the remainder columns.

b

Values indicate the direct measurement of PNCdirect and U95% C.I. that includes all recognized and evaluated sources of bias affecting PNCdirect determinations.

c

Values indicate the ratio between PNCdirect and PNCmean, expressed in percentage, and the combined standard uncertainty (uc) associated.

d

All recognized and evaluated sources of bias affecting PNC determinations were included in the parenthetical expanded uncertainty computation that corresponds to U95% C.I.