Table 4.

Ratio between PNCdirect Obtained by spICP-MS and PNCdistribution, Expressed in Percentage, for Different Commercial AuNPsab

TEM (supplier) full PSD HR-SEM Full PSD spICP-MS full PSD
30 nm PVP 88.8 ± 7.2 81.6 ± 14.4 (28.9)c 79.8 ± 6.0 (11.9)c
30 nm bPEI 92.4 ± 9.9 81.8 ± 14.9 (29.9)c 79.6 ± 7.6 (15.3)c
30 nm PEG 123.3 ± 10.6 87.7 ± 15.4 (30.7)c 86.7 ± 6.3 (12.5)c
60 nm PVP 68.8 ± 8.3 81.4 ± 9.3 (18.5)c 76.3 ± 5.4 (10.8)c
60 nm bPEI first lot 97.6 ± 13.5 82.9 ± 8.7 (17.5)c 81.0 ± 9.1 (18.3)c
60 nm bPEI second lot 94.5 ± 8.7 82.9 ± 4.0 (18.9)c 77.9 ± 4.8 (9.5)c
60 nm PEG first lot 132.6 ± 11.5 78.1 ± 8.3 (16.7)c 81.7 ± 2.1 (4.2)c
60 nm PEG second lot 113.9 ± 10.5 86.6 ± 8.9 (17.8)c 81.5 ± 7.1 (14.2)c
100 nm Citrate 114.1 ± 13.3 92.1 ± 9.5 (18.9)c 77.1 ± 6.5 (12.9)c
100 nm PVP 103.0 ± 8.4 85.0 ± 8.1 (16.1)c 73.9 ± 6.5 (13.0)c
100 nm bPEI 103.7 ± 12.2 90.2 ± 9.9 (19.8)c 87.4 ± 8.6 (17.1)c
100 nm PEG 104.9 ± 11.1 80.5 ± 5.8 (11.6)c 87.7 ± 5.4 (10.1)c
a

PNCdistribution values reported for comparison to PNCdirect were derived based on the combination of in-house Au mass fraction determination and PSD reported by TEM,33 HR-SEM,37 and spICP-MS.37

b

Values indicate the ratio between PNCdirect and PNCdistribution, expressed in percentage, and the combined standard uncertainty (uc) associated.

c

All recognized and evaluated sources of bias affecting PNC determinations were included in the parenthetical expanded uncertainty computation that corresponds to U95% C.I.