Table 7.1.

Comparison of ISO 15197:2013 and FDA blood glucose meter accuracy standards

Setting FDA (248,254) ISO 15197:2013 (255)
Home use 95% within 15% for all BG in the usable BG range
99% within 20% for all BG in the usable BG range
95% within 15% for BG ≥100 mg/dL
95% within 15 mg/dL for BG <100 mg/dL
99% in A or B region of consensus error grid
Hospital use 95% within 12% for BG ≥75 mg/dL
95% within 12 mg/dL for BG <75 mg/dL
98% within 15% for BG ≥75 mg/dL
98% within 15 mg/dL for BG <75 mg/dL

BG, blood glucose; FDA, U.S. Food and Drug Administration; ISO, International Organization for Standardization. To convert mg/dL to mmol/L, see endmemo.com/medical/unitconvert/Glucose.php.

The range of blood glucose values for which the meter has been proven accurate and will provide readings (other than low, high, or error).

Values outside of the “clinically acceptable” A and B regions are considered “outlier” readings and may be dangerous to use for therapeutic decisions (256).