TABLE 2.

Accuracy of fetal heart rate measurements when comparing transabdominal fetal electrocardiography (TAfECG) and Doppler signal with the gold standard of fetal scalp electrode (FSE). The average difference was computed as an absolute number.

TAfECG vs FSE
Average difference (bpm) SD 95% CI Limits of agreement
Lower Upper
Overall 3.5 5.3 1.88–4.12 −6.9 13.9
Static positions 1.9 3.7 0.34–1.67 −5.4 9.2
Lying down 1.8 3.7 0.33–1.67 −5.5 9.1
Sitting 1.7 4.1 0.1–1.9 −6.3 9.7
Standing 2.2 3.4 1.33–2.67 −4.5 8.9
Active positions 5.04 6.9 3.65–6.35 −8.5 18.6
Walking 3.6 4.3 2.1–3.9 −4.8 12
Birthing ball 6.5 9.5 3.98–8.02 −12.2 25,1
Doppler versus FSE
Overall 13.9 7.5 11.4–14.6 −0.8 28.6
Static positions 3.2 2.5 2.55–3.45 −1.7 8.1
Lying down 2.4 2.2 1.55–2.45 −1.9 6.7
Sitting 2.8 2.2 1.55–2.45 −1.5 7.1
Standing 4.4 3.1 3.33–4.67 −1.7 10.4
Active positions 24.7 12.4 21.3–26.7 0.8 48.6
Walking 18.4 11.7 15.5–20.5 −4.5 41.3
Birthing ball 31 13.1 28.1–34 5.3 56.7

Abbreviations: bpm, beats per minute; CI, confidence interval; SD, standard deviation.