Table 8.
Parameter | GC-MS/MS |
---|---|
Injection mode | Hot-splitless; MMI injection mode |
Injection volume | 2 μL |
Inlet temperature | 280 °C |
Carrier gas | He, constant flow 1.00 mL min−1 (column 2 = 1.20 mL min−1) |
Detector temperature | - |
Makeup gas | - |
Oven program | 70 °C for 2 min 25 °C/min to 150 °C for 0 min; 3 °C/min to 200 °C for 0 min; 8 °C/min to 280 °C for 10 min hold time |
MS transfer line temperature | 280 °C |
Backflush settings | 5 min during post-run / 310 °C |
Aux EPC pressure | ~50 psi |
Inlet pressure | ~2 psi |
Column pressure | ~3 psi |
Electron energy | 70 eV |
MS1 and MS2 resolution | Wide |
Collision cell | 1.5 mL min−1 N2 and 2.25 mL min−1 He |
Source temperature | 300 °C |
Quad temperatures | 150 °C |