Skip to main content
. 2023 Jan 3;28(1):429. doi: 10.3390/molecules28010429

Table 8.

GC-MS/MS method conditions.

Parameter GC-MS/MS
Injection mode Hot-splitless; MMI injection mode
Injection volume 2 μL
Inlet temperature 280 °C
Carrier gas He, constant flow 1.00 mL min−1
(column 2 = 1.20 mL min−1)
Detector temperature -
Makeup gas -
Oven program 70 °C for 2 min 25 °C/min to 150 °C for 0 min; 3 °C/min to 200 °C for 0 min; 8 °C/min to 280 °C for 10 min hold time
MS transfer line temperature 280 °C
Backflush settings 5 min during post-run / 310 °C
Aux EPC pressure ~50 psi
Inlet pressure ~2 psi
Column pressure ~3 psi
Electron energy 70 eV
MS1 and MS2 resolution Wide
Collision cell 1.5 mL min−1 N2 and 2.25 mL min−1 He
Source temperature 300 °C
Quad temperatures 150 °C