Skip to main content
. 2023 Jan 9;23:5. doi: 10.1186/s12880-023-00962-2

Table 3.

The results of subjective image quality analysis

ACL MCL MM LM Bone Noise Artifact Overall
Reader 1
 1NSA-DLR 4.14 ± 0.35 4.90 ± 0.29 4.95 ± 0.21 5.00 ± 0.00 4.00 ± 0.00 3.71 ± 0.45 4.81 ± 0.39 3.76 ± 0.43
 4NSA 3.76 ± 0.43 4.23 ± 0.68 3.95 ± 0.79 4.29 ± 0.88 3.86 ± 0.35 3.05 ± 0.21 4.42 ± 1.14 3.00 ± 0.69
 p value 0.03 0.001 < 0.001 0.005 0.15 < 0.001 0.14 0.003
Reader 2
 1NSA-DLR 4.19 ± 0.39 4.52 ± 0.50 4.48 ± 0.50 4.48 ± 0.50 4.00 ± 0.00 3.95 ± 0.21 5.00 ± 0.00 3.95 ± 0.37
 4NSA 3.71 ± 0.63 3.81 ± 0.59 3.86 ± 0.77 3.95 ± 0.72 3.81 ± 0.39 3.05 ± 0.37 4.52 ± 1.17 2.90 ± 0.61
 p value 0.02 0.001 0.008 0.02 0.07 < 0.001 0.17 < 0.001
Reader 3
 1NSA-DLR 4.19 ± 0.39 4.62 ± 0.49 4.57 ± 0.49 4.52 ± 0.50 4.05 ± 0.21 3.95 ± 0.21 4.86 ± 0.35 3.95 ± 0.21
 4NSA 3.90 ± 0.53 4.14 ± 0.77 3.86 ± 0.94 4.00 ± 0.82 3.90 ± 0.43 3.43 ± 0.58 4.43 ± 1.00 3.29 ± 0.76
 p value 0.04 0.03 0.01 0.02 0.23 0.001 0.10 < 0.001

All variables are expressed as the mean ± standard deviation. p values that indicate significant differences are shown in bold