Table 3.

Overall deviation of fitness tracker heart rate measurements and the clinical gold standard.


Philips Apple Fitbit Garmin Withings
Valid measurement points, n 333 327 308 332 330
Failed measurements, n 3 9 28 4 6
CDRa (%) 0.89 2.67 8.33 1.19 1.79
MAEb 0.92 1.59 2.31 2.47 1.71
MAPEc (%) 1 2 4 3 2
Bias (95% CI) –0.25 (–0.42 to –0.08) 0.36 (0.09 to 0.63) 0.77 (0.28 to 1.26) –1.21 (–1.65 to –0.77) 0.05 (–0.28 to 0.40)

aCDR: cumulative dropout rate.

bMAE: median absolute error.

cMAPE: mean absolute percentage error.