Assessment of linear correlation.

Indicator TPOa Apple Fitbit Garmin Withings
PCCb, r
(95% CI)
0.99
(0.99-0.99)
0.98 (0.98-0.99) 0.95 (0.93-0.96) 0.96 (0.95-0.96) 0.97 (0.97-0.98)
RSSc 804 1830 5133 4757 3106
P value (C test) 5.7 × 10–317 2.15 × 10–247 2.4 × 10–153 1.26 × 10–177 2.78 × 10–212
Slope 1.01 0.95 0.88 0.89 0.95
Shift –0.72 3.81 9.49 6.68 3.72
CCCd, rc (95% CI) 0.99 (0.99-0.99) 0.98 (0.98-0.99) 0.94 (0.93-0.95) 0.95 (0.94-0.96) 0.97 (0.97-0.98)

aTPO: transmissive pulse oximetry.

bPCC: Pearson correlation coefficient.

cRSS: residual sum of squares.

dCCC: Lin concordance coefficient.