Skip to main content
. 2023 Jan 4;9:1066851. doi: 10.3389/fvets.2022.1066851

Table 4.

Validation parameters of the ICP-MS method aimed at determining trace element and toxic metal concentrations in whole blood.

Element Limit of detection (ng/g) Limits of quantification (ng/g) Maximum linear concentrations Recovery (%) Precision coefficient of variation (%)
Fe 40 130 250 99 5.5
Cu 1.5 5 250 89 5.3
Mn 0.4 1.3 250 97 5.5
Zn 150 500 250 97 6.0
Se 1 3 250 95 5.2
Cr 1.5 5 250 116 8.8
Pb 0.4 1.3 250 97 2.0
Hg 0.04 0.13 10 93 5.2
Cd 0.02 0.07 250 108 12
As 0.2 0.7 250 98 6.1