Skip to main content
. 2022 Nov 8;47(3):256–261. doi: 10.4103/jmp.jmp_153_21

Table 2.

The values of percentage variation in surface doses measured to a depth of 5 mm by a Markus chamber and those of the treatment planning system for intensity-modulated radiotherapy plans

Depth (mm) Percentage variation with respect to Dmax
Percentage variation with respect to Dmean
Mean±SD Range Mean±SD Range
1 −0.15±5.46 −11.27-9.05 21.78±3.99 15.84-28.14
2 20.02±7.42 8.62-33.83 26.31±7.86 14.75-41.2
3 11.12±4.18 2.10-18.69 17.79±3.35 11.54-23.31
4 9.47±2.91 3.24-14.82 14.81±2.20 9.37-19.08
5 1.80±3.93 −4.61-10.30 7.75±3.10 1.40-13.32

SD: Standard deviation