Table 2.

Analyses of Capillary Refill Time measurements at different sites and position. For each comparison we report the median difference (non-parametric test for paired samples), the agreements (bias and standard deviation with limits of agreement [LoA]), and the Pearson correlation coefficient (in italic)

Standard Finger Elevated Finger Ear Semi-recumbent Ear Supine
Standard Finger p = 0.03 p = 0.52 p < 0.001
Bias -0.07 ± 0.30 LoA -0.61; 0.47 Bias 0.02 ± 0.18 LoA -0.33; 0.37 Bias 0.22 ± 0.40 LoA -0.56; 1.0
r = 0.67 [0.53;0.78]; p < 0.001 r = 0.28 [0.07;0.47]; p = 0.01 r = 0.12 [-0.09;0.33]; p = 0.27
Elevated Finger p = 0.03 p = 0.12 p < 0.001
Bias -0.07 ± 0.30 LoA -0.61; 0.47 Bias 0.12 (SD 0.33) LoA -0.52; 0.76 Bias 0.29 ± 0.39 LoA -0.47; 1.05
r = 0.67 [0.53;0.78]; p < 0.001 r = 0.37 [0.17;0.54]; p < 0.001 r = 0.08 [-0.13;0.29]; p = 0.46
Ear Semi-recumbent p = 0.52 p = 0.12 p < 0.01
Bias 0.02 ± 0.18 LoA -0.33; 0.37 Bias 0.12 ± 0.33 LoA -0.52; 0.76 Bias -0.17 ± 0.26 LoA -0.67; 0.33
r = 0.28 [0.07;0.47]; p = 0.01 r = 0.37 [0.17;0.54]; p < 0.001 r = 0.40 [0.20;0.57]; p < 0.001
Ear Supine p < 0.001 p < 0.001 p < 0.01
Bias 0.22 ± 0.40 LoA -0.56; 1.0 Bias 0.29 ± 0.39 LoA -0.47; 1.05 Bias -0.17 ± 0.26 LoA -0.67; 0.33
r = 0.12 [-0.09;0.33]; p = 0.27 r = 0.08 [-0.13;0.29]; p = 0.46 r = 0.40 [0.20;0.57]; p < 0.001