Table 2.
Analyses of Capillary Refill Time measurements at different sites and position. For each comparison we report the median difference (non-parametric test for paired samples), the agreements (bias and standard deviation with limits of agreement [LoA]), and the Pearson correlation coefficient (in italic)
Standard Finger | Elevated Finger | Ear Semi-recumbent | Ear Supine | |
---|---|---|---|---|
Standard Finger | p = 0.03 | p = 0.52 | p < 0.001 | |
Bias -0.07 ± 0.30 LoA -0.61; 0.47 | Bias 0.02 ± 0.18 LoA -0.33; 0.37 | Bias 0.22 ± 0.40 LoA -0.56; 1.0 | ||
r = 0.67 [0.53;0.78]; p < 0.001 | r = 0.28 [0.07;0.47]; p = 0.01 | r = 0.12 [-0.09;0.33]; p = 0.27 | ||
Elevated Finger | p = 0.03 | p = 0.12 | p < 0.001 | |
Bias -0.07 ± 0.30 LoA -0.61; 0.47 | Bias 0.12 (SD 0.33) LoA -0.52; 0.76 | Bias 0.29 ± 0.39 LoA -0.47; 1.05 | ||
r = 0.67 [0.53;0.78]; p < 0.001 | r = 0.37 [0.17;0.54]; p < 0.001 | r = 0.08 [-0.13;0.29]; p = 0.46 | ||
Ear Semi-recumbent | p = 0.52 | p = 0.12 | p < 0.01 | |
Bias 0.02 ± 0.18 LoA -0.33; 0.37 | Bias 0.12 ± 0.33 LoA -0.52; 0.76 | Bias -0.17 ± 0.26 LoA -0.67; 0.33 | ||
r = 0.28 [0.07;0.47]; p = 0.01 | r = 0.37 [0.17;0.54]; p < 0.001 | r = 0.40 [0.20;0.57]; p < 0.001 | ||
Ear Supine | p < 0.001 | p < 0.001 | p < 0.01 | |
Bias 0.22 ± 0.40 LoA -0.56; 1.0 | Bias 0.29 ± 0.39 LoA -0.47; 1.05 | Bias -0.17 ± 0.26 LoA -0.67; 0.33 | ||
r = 0.12 [-0.09;0.33]; p = 0.27 | r = 0.08 [-0.13;0.29]; p = 0.46 | r = 0.40 [0.20;0.57]; p < 0.001 |