Skip to main content
. 2023 Feb;64(2):159–166.

Table 5.

Two bag tests of the rebreathing system.

Item on checkout sheet Description Percentage of applicable machines that passed the checkout procedure
Two Bag Test A Rebreathing system is patent. 100% (89/89)
Two Bag Test B Unidirectional valve function during manual ventilation. 100% (89/89)
Two Bag Test C APL valve functions. 100% (89/89)
Two Bag — Part D Ventilator present. 14.6% (13/89)
Two Bag — Part E Breathing system protected with reservoir bag when not in use. 0% (0/89)

Two bag test items of the checkout procedure were not included in machines without a rebreathing system (n = 11).