Table 5.
Two bag tests of the rebreathing system.
Item on checkout sheet | Description | Percentage of applicable machines that passed the checkout procedure |
---|---|---|
Two Bag Test A | Rebreathing system is patent. | 100% (89/89) |
Two Bag Test B | Unidirectional valve function during manual ventilation. | 100% (89/89) |
Two Bag Test C | APL valve functions. | 100% (89/89) |
Two Bag — Part D | Ventilator present. | 14.6% (13/89) |
Two Bag — Part E | Breathing system protected with reservoir bag when not in use. | 0% (0/89) |
Two bag test items of the checkout procedure were not included in machines without a rebreathing system (n = 11).