Skip to main content
. 2021 Dec 10;22(1):103–115. doi: 10.2463/mrms.tn.2021-0094

Table 3.

Signal intensity and loss measured from the FID measurement (DEC, from the crossband converter to the 1H-MRI system)

System Signal intensity [dBm] Total receiver gain [dB] Relative signal loss compared with A [dB]
Conventional 23Na-MRI System (A) –79.0 83.7 -
Add-on system (B) –78.6 104.4 21.1
Add-on system with DEC (C) –78.5 92.3 10.9

DEC, direct bypass electrical connection; FID, free induction decay.