Skip to main content
. 2022 Feb 26;22(1):7–25. doi: 10.2463/mrms.rev.2021-0047

Table 4.

Temperature safety limits in the IEC standard

Operating mode Maximum core temperature (°C) Maximum local tissue temperature (°C) Rise of core temperature (°C)
Normal 39 39 0.5
First-level controlled 40 40 1

For the second-level-controlled operating mode, determining the limits is the responsibility of the investigational review board. IEC, International Electrotechnical Commission.