Table 1.

Clinical applications of mucosal GES in recent years.

Time Model Position of electrodes Instrument Conclusions
2005 Healthy humans 5 cm above the pylorus A universal pulse generator (Acupulser, model A310, World
Precision Instrument, Inc., Sarasota, FL, USA):
Frequency 9 cpm; pulse width 500 ms;amplitude 5 mA
Acute RGES can result in a series of symptoms associated with dyspepsia, which is beneficial to the treatment of obesity. Optimal parameter should be determined according to the individual sensitivity to electrical stimulation.
2005 Patients with gastroparesis The junction of the antrum and the body of the stomach Enterra:
Frequency14 Hz;
amplitude5–10 mA;
pulse width 330 μs;
cycle ON 0.1–1.0 s; cycle OFF 5.0–4.0 s.
Endoscopic placement of mucosal GES electrodes are safe and effective in patients with gastroparesis.
2005 Healthy humans The greater curvature 5 cm above the pylorus A universal pulse generator (Acupulser, model A310, World
Precision Instrument, Inc., Sarasota, FL, USA):
Long-pulse:frequency 9 cpm; pulse width 500 ms; amplitude 5 mA, 10 mAor15 mA;
Pulse train:2 s ON, 3 s OFF; frequency20 Hz;
pulse width5 ms, 10 ms
or 20 ms; amplitude10
mA, 15 mA or 20 mA.
GES via the distal stomach reduces gastric accommodation and delays gastric emptying.
2006 Healthy humans The greater curvature 5–6 cm away from the gastroesophageal junction. An adjustable stimulator (model A310, World Precision Instruments, Inc., Sarasota, FL):
Amplitude5–10 mA;
pulse width 100–500 ms; frequency 9 cpm;
10 min ON, 10 min OFF.
GES using temporary mucosal electrodes decreases food intake as well as maximum intake of water,and has a tendency of delaying gastric emptying. It may have a potential application for the treatment of obesity.
2011 A man with diabetic gastroparesis The junction of the gastric body and antrum Enterra ENDOStim is the preferred method for placement of electrodes for temporary GES.
2013 Patients with gastroparesis The junction of the gastric body and antrum Enterra:
Impedance: 200–800 Ω; current: 5 mA; cycle time: 0.1 s ON, 5 s OFF
A trial of temporary, endoscopically delivered GES may be of predictive value to select patients for laparoscopic- implantation of a permanent GES device.
2015 Patients with gastroparesis The junction of the gastric body and antrum Implantable pulse generator (Medtronic 3116) GES may be an effective therapy for treating the symptoms of gastroparesis with normal gastric emptying.
2016 Patients with drug refractory, cyclic vomiting syndrome The junction of the gastric body and antrum Enterra:
Frequency, 14 Hz; intensity, 5 mA; pulse width, 330 μs; cycle ON, 0.1 s; cycle OFF, 5 s.
In a small group of drug-refractory CVS patients, treatments with temporary and permanent GES significantly reduced the severity of gastrointestinal symptoms and frequency of hospital admissions.
2018 Patients with gastroparesis 10 cm from the pylorus at the greater curvature and separated approximately 1 cm Enterra:
Frequency, 14 Hz; intensity, 5 mA; pulse width, 330 μs; cycle ON, 0.1 s; cycle OFF, 5 s.
Laparoscopic implantation of a temporary GES system predicts the outcome of permanent GES and is cost-saving.