. 2023 Jan 6;9:1099014. doi: 10.3389/fcvm.2022.1099014

TABLE 2.

Details on monitoring.

A
All ILR Smartphone ECG p
N = 116 N = 23 N = 93
Median time to first arrhythmia, days (IQR) 92 (16–233) 40 (15–681) 102 (21–232) 0.801
Median monitoring time per patient, days (IQR) 322 (148–428) 567 (40–1217) 317 (188–399) 0.045
B
Details on first arrhythmia
All ILR Smartphone p
N = 116 N = 23 N = 93
Arrhythmia occured, N (%) 33 (28) 11 (48) 22 (24) 0.021
Atrial fibrillation, N (%) 16 (14) 2 (9) 14 (15) 0.428
Supraventricular tachycardia, N (%) 14 (12) 6 (26) 8 (9) 0.021
Ventricular tachycardia, N (%) 1 (1) 1 (4) 0 0.043
Sinus node defect, N (%) 2 (2) 2 (9) 0 0.004
Atrioventricular block, N (%) 0 0 0 1.000

N, number; IQR, interquartile ranges; ILR, implantable loop recorder. Bold values represent the significant values.