Skip to main content
. 2023 Jan 11;13(2):267. doi: 10.3390/diagnostics13020267

Table 2.

Table comparing posterior Ks(D) and Kf (D), Ks axis(°), Kf axis(°), posterior astigmatism As.(D) measured with OCT Casia 2 and GalileiG6 Dual Scheimpflug Analyzer. ICC (introclass correlation coefficient) refers to the correlation between devices.

GalileiG6 ss-OCT Casia 2 Difference: GalileiG6-ss-OCT Casia 2 p * Limits of Agreement ICC (95% CI)
Lower (95% CI) Upper (95% CI)
Ks (D) (Mean ± SD) −6.56 ± 0.51 −6.33 ± 0.3 −0.23 ± 0.42 p < 0.001 −1.04 (−1.05; −1.04) 0.59 (0.58; 0.59) 0.43 (0.23; 0.59)
Kf (D) (Mean ± SD) −6.13 ± 0.36 −6.07 ± 0.26 −0.06 ± 0.26 p < 0.001 −0.57 (−0.57; −0.57) 0.44 (0.44; 0.44) 0.65 (0.55; 0.74)
Ks axis (°) (Mean ± SD) 79.29 ± 32.08 87.04 ± 17.6 −7.75 ± 31.31 p = 0.021 −69.11 (−102.79; −35.43) 53.61 (19.93; 87.29) 0.26 (0.1; 0.4)
Kf axis (°) (Mean ± SD) 104.43 ± 67.06 96.1 ± 80.3 8.33 ± 77.45 p = 0.619 −143.48 (−349.66; 62.7) 160.14 (−46.04; 366.32) 0.45 (0.31; 0.57)
Ast. (D) (Mean ± SD) 0.45 ± 0.36 0.27 ± 0.13 0.49 ± 0.36 p < 0.001 −0.92 (−0.79; −1.05) 0.56 (0.69; 0.44) 0 (−0.13; 0.14)

* Paired Wilcoxon test.