Table 3.

Table comparing central corneal thickness CCT (um), TCT (um) measured with OCT Casia 2 and GalileiG6 Dual Scheimpflug Analyzer. ICC (introclass correlation coefficient) refers to the correlation between devices.

GalileiG6 ss-OCT Casia 2 Difference: GalileiG6-ss-OCT Casia 2 p * Limits of Agreement ICC (95% CI)
Lower (95% CI) Upper (95% CI)
CCT (um) (Mean ± SD) 566.71 ± 39.31 547.83 ± 34.28 18.88 ± 22.62 p < 0.001 −25.45 (−43.04; −7.87) 63.21 (45.63; 80.8) 0.72 (0.37; 0.85)
TCT (um) (Mean ± SD) 551.67 ± 39.77 539.38 ± 34.13 12.29 ± 25.27 p < 0.001 −37.25 (−59.2; −15.29) 61.83 (39.87; 83.78) 0.73 (0.58; 0.82)

* Paired Wilcoxon test.