Independent monitor unit verification (MUV) methods for the dynamic beam‐flattening (DBF) technique have not been established. The purpose of this study was to clarify whether MU values for the DBF technique can be calculated using in‐air and in‐water output ratios (S c and S cp). S c and S cp were measured in the DBF mode, and the phantom scatter factor (S p) was calculated. The difference between calculated and planned MUs with square and rectangle fields and clinical plans for different treatment sites was also evaluated. S c values for the 4 × 4 to 24 × 24 cm2 fields of the distal multi‐leaf collimator (MLC) layer at 2‐cm intervals were 0.887, 0.815, 0.715, 0.716, 0.611, 0.612, 0.511, 0.373, 0.374, 0.375, and 0.374, respectively. No collimator exchange effect was observed. S c also depends slightly on the field size of the distal MLC layer. If the distal‐MLC‐layered field size was less than 20% of the corresponding MLC sequence size in the proximal MLC layer, S c was affected by >1%, which was compensated using a correction factor (CF). S p increased as the field sizes of the MLC sequence and distal MLC leaves increased. MUs calculated using measured S c, S p, and CF for square and rectangle fields agreed with planned MUs within ±1.2%. A larger difference (−1.5%) between calculated and planned MUs was observed for clinical plans, whereas differences in MUs were within 2 MU for most fields (56 out of 64 fields). MU calculation for the DBF technique can be performed with S c, S p, and CF for independent MUV.
Keywords: dynamic beam flattening, halcyon, in‐air output, monitor unit verification
1. INTRODUCTION
Independent monitor unit verification (MUV) is an efficient quality control (QC) method for detecting pretreatment errors affecting the treatment planning system dose calculations. 1 A report of Task Group 114 of the American Association of Physicists in Medicine concluded that MUV remains an important element of the radiation therapy QA/QC program that ensures a safe and accurate patient treatment in modern radiation treatment plan, which uses volumetric imaging, improved computation algorithm, and heterogeneity correction. 1 MU calculation for photon beams may be performed using dose per monitor unit at the point of interest, in‐air output ratio (S c), phantom scatter factor (S p), and tissue phantom ratio (TPR). These factors depend on the field size because of variation in the scattered radiation originating from the collimator head or the phantom, especially S c depends on the field size at the virtual source plane that can be seen at the point of interest (i.e., detector's eye view). 1 , 2 , 3 , 4 , 5 The dosimetric function values for irregular fields are approximated by using the equivalent field size, which is defined as the square field with the same depth–dose characteristics as the irregular field. 6 The equivalent field sizes are determined from the blocked field area of the jaws for S c and multi‐leaf collimator (MLC) for S p. 7 , 8 , 9 , 10 , 11 , 12
A new radiotherapy delivery system, the Halcyon (Varian Medical Systems, Palo Alto), was introduced into clinical practice. 13 Halcyon is designed for faster delivery and higher throughput using a single 6‐MV flattening‐filter‐free (FFF) beam equipped with a dual‐layer MLC and a maximum field size of 28 × 28 cm2 compared with a conventional treatment unit. 14 , 15 In Halcyon version 2.0, dynamic beam flattening (DBF) is used to mimic flattened beams for conventional radiation therapy treatments. 16 In this technique, the distal layer of the MLC leaves is used to define a static MLC aperture, and the proximal layer is used to flatten the FFF beam profile with a sliding‐window leaf motion. Halcyon can be used to deliver all types of treatments; however, the independent verification of MU calculations has not been established for DBF plans. Different treatment planning systems cannot be used to verify MU values for DBF plans because only Eclipse (Varian Medical Systems, Palo Alto) supports the creation of plans for the Halcyon. Therefore, DBF plans must be verified with measurements obtained for patient‐specific QC, similar to intensity‐modulated radiation therapy or volumetric modulated arc therapy pretreatment verification. 17
The purpose of this study was to clarify whether MU for the DBF technique can be calculated with S c and S p for MUV and to propose a simple MUV method for DBF.
2. METHODS
2.1. Dynamic beam flattening (DBF)
The Halcyon MLC system features a unique stacked‐and‐staggered dual‐layered design consisting of a distal and a proximal layer. The distal layer, placed further from the source, comprises two banks with 28 leaves each. The proximal layer, placed closer to the source, comprises two banks with 29 leaves each. 18 The layers are located with a 0.5‐cm offset in the direction perpendicular to the traveling direction of the leaves. 19 The width of all leaves is 1 cm at the isocenter plane.
In the DBF technique, the proximal layer of the MLC leaves moves according to predefined MLC sequences to compensate the FFF beam profile. The MLC sequence has predefined sweeping pattern with numbers of control points. Predefined MLC sequences are for the following field sizes (X × Y): 5 × 5, 7 × 7, 11 × 11, 15 × 15, 19 × 19, 23 × 23, and 24 × 28 cm2, where X and Y are the directions parallel and perpendicular to the traveling direction of the leaves, respectively (Figure 1). Those are only available predefined ones from the system. Two MLC sequences for the 24 × 28 cm2 field are provided: with and without shielded corners, which are used to provide better protection against leakage at the closing positions of the leaf pairs in the distal MLC layer. When user shapes a field by the distal MLC layer, the MLC sequence for the proximal layer is automatically selected by the system. The system selects the smallest possible MLC sequence for the aperture defined by the user. DBF aims for a flat profile at a depth of 10 cm, where flatness is defined as a deviation of no more than ±3% from the dose at the central axis for 80% of the central field of the 10 × 10 to 22 × 22 cm2 fields. 13 , 20 The flatness for the field sizes smaller than 10 × 10 cm2 is not defined in manufacturer's reference guide. 20
FIGURE 1.

Dynamic beam flattening multi‐leaf collimator (MLC) sequence boundaries in beam's‐eye‐view and relationship between distal‐MLC‐layered field size and MLC sequence selected by the system automatically
2.2. Measurement of in‐air output ratio (S c)
In‐air outputs for the distal‐MLC‐layered field sizes of 4 × 4 to 24 × 24 cm2 at 2‐cm intervals were measured at isocenter using an 0.6 cm3 ionization chamber (PTW, Freiburg) inserted into a cylindrical mini‐phantom (Qualita, Nagano) with 4 cm in diameter and 20 cm in length. The ion chamber was placed 10 cm below the surface of the mini‐phantom. As MLC sequences of the 23 × 23 and 24 × 28 cm2 fields with shielded corners are not used for a square field (Figure 1), in‐air outputs for the 20 × 16 and 12 × 24 cm2 fields of the distal MLC layer were measured instead of them. To calculate S c, all in‐air outputs were normalized to the in‐air output of the 10 × 10 cm2 static field produced in the FFF mode. All measurements were repeated three times with 100 MU, and the values were averaged.
In addition, in‐air outputs for the 14 × 4 , 4 × 14 , 18 × 4, and 18 × 4 cm2 fields were measured to determine the magnitude of the collimator exchange effect. 21
In‐air outputs for rectangle fields were measured to determine the effect of the field size of the distal MLC layer on S c. The X side of the field was fixed at 4 cm for all rectangle fields, and the Y side of the field was changed from 8 to 28 cm at 2‐cm intervals. Because the diameter of the mini‐phantom used for this study was 4 cm, lateral equilibrium was not achieved for smaller fields (<4 × 4 cm2).
2.3. Measurement of in‐water output ratio (S cp) and calculation of a phantom scatter factor (S p)
In‐water output measurements were carried out using the same 0.6‐cm3 ionization chamber placed at a depth of 10 cm in a 35 × 35 × 40 cm3 water phantom (Qualita, Nagano) with a source‐to‐surface distance (SSD) of 90 cm for the same field sizes as the S c measurements. This measurement depth was determined to meet the calibration depth for the Halcyon in our center. To calculate S cp, all in‐water outputs were normalized to the in‐water output of the 10 × 10 cm2 static field produced in the FFF mode. Thereafter, S p was calculated using S cp and S c for each field size according to the following equation:
| (1) |
2.4. Comparison of calculated and planned monitor units (MUs)
S c and S p obtained in this study were evaluated by comparing the calculated and planned MUs. The planned MUs were obtained from Eclipse treatment planning system version 15.6 (Varian Medical Systems, Palo Alto). The MUs for square and rectangle fields were calculated to deliver 100 cGy to the isocenter at a depth of 10 cm with an SSD of 90 cm in a water‐cube phantom. The dose calculation algorithm, Anisotropic Analytical Algorithm (AAA) version 15.6.06 (Varian Medical Systems, Palo Alto), was used without heterogeneity correction. For MUV, the determination of the field size for S p was based on the field size of the distal MLC leaves. The method of Sterling et al. 22 was used to calculate the equivalent square‐field size. Determination of S c was based on the DBF MLC sequence used for the field. The equation for MU is as follows:
| (2) |
where D 0 is the dose per MU at the normalization depth (10 cm) for 10 × 10 cm2, D is the absorbed dose at the point of interest, and r d is the equivalent square of the treatment aperture defined by collimating devices projected to the plane normal to the central axis containing the point of calculation at a depth d. In this study, the point of interest and calculation depth were 10 cm. The TPR was 1.0, and the effect of TPR errors on the accuracy of S c and S p was minimized. CF is a correction factor for S c, introduced to consider the effect of the field size of the distal MLC layer on S c. The details of CF are described in the following subsection. The standard SSD (SSD0) is the distance along the central axis from the physical source to the patient/phantom surface under normalization conditions, and d 0 is the normalization depth. The source‐to‐axis distance (SAD) is the distance between the X‐ray physical source position and the isocenter. The values of SSD0 + d 0 and SAD were equal in this study.
2.5. Correction factor for in‐air output on the Halcyon linac
The machine output can deviate from the predicted output when the area of the field shaped by the MLC is less than ∼50% of the original field area. 23 Nara et al. introduced a new method to correct S c for fields shaped by MLC in which the CF is determined by the ratio of S c for an MLC square field to S c for a jaw square field. 24 The Halcyon does not have a jaw, but larger differences between the proximally and distally defined field areas may provide different S c. We also attempted to correct S c for rectangle fields in the Halcyon by referring to the method of Nara et al. The ratio of S c for a rectangle field to S c for a square field was calculated using the results of S c measurements. All field area were defined at the isocenter distance. These ratios were plotted against the ratio of the field area of the MLC sequence defined by the proximal leaves to the field area of the distal MLC leaves, and a regression curve was determined for this plot. The field area of the MLC sequence was defined based on the sequence boundaries shown in Figure 1. Thus, an S c CF can be applied according to the ratio of the field area of the MLC sequence to the field area of the distal MLC leaves.
2.6. Comparison of calculated and planned monitor units (MUs) with clinical plans
For further comparison, the difference between calculated and planned MUs in irregular‐shaped fields was evaluated. A total of 64 fields was collected from 19 clinical plans for patients with metastatic brain tumor (2 plans), esophagus cancer (2 plans), lung cancer (3 plans), breast cancer (3 plans), pancreas and liver cancer (3 plans), pelvic cancer (4 plans), and metastatic spinal tumor (2 plans). All plans were created with two to five fields, and AAA was used to calculate the dose distribution. Prescription doses were 2–3 Gy at the isocenter. To eliminate the error caused by patient geometry, all fields were recalculated with a water‐cube phantom. The calculation point was placed at a depth of 10 cm. MUV was performed using the same geometry without homogeneity correction. The method of Onai et al. 25 was used to calculate the equivalent square‐field size for the determination of S p. Their method is simpler than Clarkson's method 26 to estimate S p for irregular‐shaped fields at the isocenter.
2.7. Difference between MUV and portal dosimetry results for clinical plans
Our MUV was performed for some plans of the spine and whole brain, which were evaluated using portal dosimetry (PD; Varian Medical Systems, Palo Alto), 27 , 28 and then delivered to patients in the clinic. The calculated and measured fluences were compared in the PD workspace incorporated in Eclipse. The results of the gamma evaluation with a dose difference of 3%, a distance‐to‐agreement of 2 mm, 29 and a low‐dose threshold of 10% were compared with the results of MUV. The MUs for fields were calculated using patient geometry and depths without inhomogeneity correction, and the conditions were more advanced than those of this study.
3. RESULTS
3.1. Measurement of in‐air output ratio (S c)
S c values were 0.887, 0.815, 0.715, 0.716, 0.611, 0.612, 0.511, 0.373, 0.374, 0.375, and 0.374 for the 4 × 4 to 24 × 24 cm2 fields of the distal MLC layer at 2‐cm intervals, respectively (Figure 2). S c values were 0.411 and 0.434 for the 20 × 16 and 12 × 24 cm2 fields of the distal MLC leaves, respectively.
FIGURE 2.

Ratio of in‐air output (S c) for square fields with size of multi‐leaf collimator (MLC) sequence used. MLC sequences for the 23 × 23 and 24 × 28 cm2 fields with shielded corners were not used for the square fields. To obtain S c in these MLC sequence fields, 20 × 16 and 12 × 24 cm2 fields were used, and the values of S c were 0.411 and 0.433, respectively. Using the same MLC sequence (shown in this figure as different colors), the same value of S c was observed even if the field size of the distal MLC leaves was different.
S c values for the 14 × 4 and 4 × 14 cm2 fields were 0.605 and 0.606, respectively. S c values for the 18 × 4 and 4 × 18 cm2 fields were 0.503 and 0.504, respectively. The collimator exchange effect was very insignificant.
To investigate the effect of the field size of the distal MLC layer on S c, measurements were conducted with one side of each field (X side) set to 4 cm. Figure 3 shows the ratio of S c for a rectangle field to S c for a square field. There were 30 cases in Figure 3, and all values of measured field sizes in this study were plotted. S c changed by more than −1% when the ratio of the rectangle‐field area to square‐field area was less than 20%. A polynomial regression curve was added to the plot in Figure 3, and a CF for S c depending on the ratio of the proximal and distal field areas was determined.
FIGURE 3.

Correction factor of S c as a function of the ratio of multi‐leaf collimator (MLC) sequence size to field size of distal MLC leaves. There were 30 cases in this figure, and all values of measured field sizes in this study were plotted: 11 square fields, 19 rectangular fields.
3.2. Measurement of in‐water output ratio (S cp) and calculation of a phantom scatter factor (S p)
Parts (a) and (b) of Figure 4 show the results for S cp and S p, respectively. S cp decreased as the field size of the MLC sequence increased. S cp slightly increased as the field size of the distal MLC leaves increased when the same MLC sequence was used for beam flattening. S p increased as the field sizes of the MLC sequence and distal MLC leaves increased.
FIGURE 4.

(a) Ratio of in‐water output (S cp) for square fields with size of multi‐leaf collimator (MLC) sequence used and (b) phantom scatter factor (S p) for square fields with size of MLC sequence used
3.3. Comparison of calculated and planned monitor units (MUs)
Calculated and planned MUs for square and rectangle fields, summarized in Table 1, were compared. The largest difference (−1.2%) between the calculated and planned MUs was observed for the 6 × 28 and 8 × 28 cm2 fields. Calculated and planned MUs for most field sizes agreed (within ±1.0%).
TABLE 1.
Difference (%) of calculated MUs relative to planned MUs from Eclipse version 15.6 for square and rectangle fields at a depth of 10 cm with a source‐to‐surface distance of 90 cm
| Field size of X direction (cm) | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Field size of Y direction (cm) | 4.0 | 6.0 | 8.0 | 10.0 | 12.0 | 14.0 | 16.0 | 18.0 | 20.0 | 22.0 | 24.0 |
| 4.0 | 0.0 | −0.4 | 0.1 | −0.1 | −0.1 | −0.3 | −0.3 | −0.5 | −0.6 | −0.7 | −0.6 |
| 6.0 | −0.3 | 0.0 | 0.1 | 0.0 | −0.2 | −0.3 | −0.5 | −0.7 | −0.8 | −1.0 | −1.0 |
| 8.0 | 0.2 | 0.1 | 0.0 | 0.2 | −0.2 | −0.3 | −0.5 | −0.6 | −0.8 | −0.9 | −0.9 |
| 10.0 | 0.0 | 0.0 | 0.2 | 0.5 | 0.0 | 0.0 | −0.4 | −0.4 | −0.6 | −0.7 | −0.7 |
| 12.0 | −0.0 | −0.2 | −0.2 | 0.0 | 0.1 | 0.2 | −0.2 | −0.1 | −0.4 | −0.4 | −0.4 |
| 14.0 | −0.4 | −0.2 | −0.3 | −0.3 | 0.0 | 0.5 | 0.0 | 0.1 | −0.1 | −0.1 | −0.1 |
| 16.0 | −0.2 | −0.5 | −0.5 | −0.4 | −0.2 | 0.0 | 0.2 | 0.4 | 0.1 | 0.0 | −0.1 |
| 18.0 | −0.4 | −0.6 | −0.6 | −0.5 | −0.2 | 0.1 | 0.4 | −0.2 | 0.1 | 0.1 | 0.1 |
| 20.0 | −0.5 | −0.8 | −0.8 | −0.7 | −0.4 | −0.2 | 0.0 | 0.1 | 0.1 | 0.2 | 0.2 |
| 22.0 | −0.7 | −1.0 | −1.0 | −0.8 | −0.5 | −0.2 | −0.1 | 0.1 | 0.2 | 0.2 | 0.2 |
| 24.0 | −0.6 | −0.9 | −0.9 | −0.8 | −0.5 | −0.3 | −0.1 | 0.0 | 0.2 | 0.2 | 0.1 |
| 26.0 | −0.7 | −1.1 | −1.1 | −0.9 | −0.5 | −0.4 | −0.2 | 0.0 | 0.1 | 0.1 | 0.0 |
| 28.0 | −0.8 | −1.2 | −1.2 | −1.0 | −0.6 | −0.5 | −0.2 | 0.0 | 0.0 | 0.0 | −0.3 |
Figure 5 shows the results of comparing the calculated and planned MUs for irregular‐shaped fields by case. The largest difference (−1.5%) between calculated and planned MUs was observed for pelvis case, and the smallest difference (0.1%) was observed for lung case. Differences in MUs were within 2 MU for most fields (56 out of 64 fields). The largest difference between calculated and planned MUs was 3.4 MU for brain case. Calculated MUs were systematically lower than planned MUs for most fields.
FIGURE 5.

Difference (%) of calculated MUs relative to planned MUs from Eclipse version 15.6 for different treatment sites in clinical plans
3.4. Difference between MUV and portal dosimetry results for clinical plans
The results of MUV were 1.46% and 0.35% for two fields in the spine plan and 1.51% and −1.05% for two fields in the whole brain plan compared with planned MUs for the fields, whereas the gamma pass rates were >99% for all fields.
4. DISCUSSION
The results of this study demonstrate that S c can be determined from the DBF MLC sequence, and S p can be determined from the field size of the distal MLC leaves in the DBF mode. The MU for a DBF field can be calculated to planned MUs within ±2% using the look‐up tables as independent MUV for conventional linear accelerators.
S c was dominated by MLC sequence field size in Figure 1. In a conventional treatment machine with traditional upper and lower jaws, S c is determined on the basis of the field size of the collimator jaws. However, if the field size is smaller (<0.5, i.e., smaller than half of the field area) with an MLC than with collimator jaws, the value of S c may change. 23 In the Halcyon, the proximal MLC layer plays a role like the collimator jaws to determine S c value for the field size. To calculate MUs, S c was selected on the basis of the MLC sequence used for the field, and CF was applied to account for the effect of the field size of the distal MLC leaves. The field size of the distal MLC leaves affected S c, which depended on the ratio of field area of the DBF MLC sequence to the field area of the distal MLC leaves. Nara et al. attempted to calculate accurate S c values using CFs for irregular‐shaped fields with MLC. Therefore, we applied CF to correct S c and eliminate the effect of the field size of the distal MLC layer. S p increased as the field sizes of the MLC sequence and distal MLC leaves increased. For rectangular fields, the equivalent square field converted from the field size of the distal MLC leaves can be used successfully, even if the proximal MLC leaves move according to the preconfigured MLC sequence during irradiation in the DBF mode. With square and rectangle fields, the difference between calculated and planned MUs was less than 1.5% in this study. With irregular‐shaped fields in clinical plans, the difference between calculated and planned MUs was less than 2%, which was slightly larger than those of square and rectangle fields, and calculated MUs were systematically lower than planned MUs. These discrepancies likely originated in the accuracy of the calculated equivalent square field for the determination of S p. Moreover, Kim et al. found that measured dosimetric leaf gap sizes for each layer were different compared with defined value in the TPS. 19 The suboptimal leaf end modeling may affect the results systematically. However, the differences were small. Some previous studies present guidelines for action levels for disagreement between planned and calculated MUs, depending on treatment site, treatment technique, patient geometry, and whether corrections for tissue heterogeneities are used. 1 , 30 , 31 They concluded action levels of 3%–5% were clinically realistic based on collected data from multi‐institutions or the collective experience and expectations of the task group members. The results of this study indicate that our method is sufficiently accurate for MUV of DBF fields in clinical practice. Verifying MU without measurements for patient plans leads to improved workflow in the clinic, thus reducing the waiting time for radiotherapy. Plans for the Halcyon can only be created using Eclipse, and third‐party treatment planning systems cannot be used for MUV. Additionally, all Halcyon treatment plans use dynamic MLC sequences even if the plan is used for conventional radiotherapy. As things stand, the dose in all plans should be measured for patient‐specific QC. Our results support the establishment of independent MUV methods for DBF fields on the Halcyon to detect pretreatment errors affecting the TPS dose calculations.
In this study, the collimator exchange effect was very small for the dual‐layered MLC in the Halcyon. For linear accelerators equipped with conventional jaws, because the upper collimator jaws are closer to the source than the lower jaws, the same opening size made by the upper and lower jaws in the beam's eye view are different when viewed from the perspective of calculations, and thus, the collimator backscattered radiation, which contributes to the monitor unit signal, from each jaw is different. In the Halcyon, the same MLC sequence is used in the proximal MLC layer even if the field size of the distal MLC layer changes, and thus the collimator backscattered radiation from each layer is the same. In addition, FFF beam related to the reduction of head scatter would be also the reason. 32
In addition, Our MUV method may be applied instead of PD verification for DBF plan in the Halcyon and may lead to the independency of the verification process because the PD is a system provided by the manufacturer.
There are some limitations in this study. Calculation point was the isocenter for all fields, and homogeneity correction was not used. Difference between calculated and planned MUs might indicate a different tendency when a calculation point at off‐axis is used or heterogeneity tissues is considered. To calculate MUs in clinical practice, TPR is required for different calculation depth, but it was out of scope in this study. We further investigate them in the future study. In order to validate whether our method is sufficiently accurate for MUV of DBF fields in clinical practice, the following three steps would be required: (1) factor model design and commissioning using measurements in regular fields, (2) characterization of the differences in MU calculations between the factor model and the TPS, and (3) defining criteria for a clinical MUV procedure based on this characterization/analysis. In this study, we focused on steps (2) and (3), but step (1) was not confirmed with adequately validation set. The beam data to be used for dose calculation for the Halcyon are provided from manufacturer (i.e., representative beam data), and users cannot do any modifications. We assumed that the beam data and measurement in regular fields meet within clinical criteria.
5. CONCLUSION
MU calculation for the DBF technique can be performed with S c, S p, and CF for independent MUV. This simple method can be used to verify DBF plans, which may reduce the frequency of measurement for patient‐specific QC.
AUTHOR CONTRIBUTIONS
Kazuki Kubo and Hajime Monzen conceived the idea of the study. Kazuki Kubo, Mikoto Tamura, Kenji Matsumoto, and Masakazu Otsuka significantly contributed to data analysis and interpretation. All authors reviewed the manuscript draft and revised it critically on intellectual content. All authors approved the final version of the manuscript to be published.
CONFLICTS OF INTEREST
The authors report no conflicts of interest. The authors are responsible for the content and writing of the paper.
ACKNOWLEDGMENTS
We thank Edanz (https://jp.edanz.com/ac) for editing a draft of this manuscript.
Kubo K, Tamura M, Matsumoto K, Otsuka M, Monzen H. Independent monitor unit verification for dynamic flattened beam plans on the Halcyon linac. J Appl Clin Med Phys. 2023;24:e13807. 10.1002/acm2.13807
REFERENCES
- 1. Stern RL, Heaton R, Fraser MW, et al. Verification of monitor unit calculations for non‐IMRT clinical radiotherapy: report of AAPM Task Group 114. Med Phys. 2011;38(1):504‐530. 10.1118/1.3521473 [DOI] [PubMed] [Google Scholar]
- 2. Karlsson M, Ahnesjö A, Georg D, Nyholm Z. Independent Dose Calculations, Concepts and Models. ESTRO Booklet Series No. 10. ESTRO; 2010. [Google Scholar]
- 3. Gibbons JP, Antolak JA, Followill DS, et al. Monitor unit calculations for external photon and electron beams: report of the AAPM Therapy Physics Committee Task Group No. 71. Med Phys. 2014;41(3):031501. 10.1118/1.4864244 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Kim S, Palta JR, Zhu TC. A generalized solution for the calculation of in‐air output factors in irregular fields. Med Phys. 1998;25(9):1692‐1701. 10.1118/1.598350 [DOI] [PubMed] [Google Scholar]
- 5. Yang Y, Xing L, Boyer AL, Song Y, Hu Y. A three‐source model for the calculation of head scatter factors. Med Phys. 2002;29(9):2024‐2033. 10.1118/1.1500767 [DOI] [PubMed] [Google Scholar]
- 6. Day MJ, Aird EG. The equivalent‐field method for dose determinations in rectangular fields. Br J Radiol Suppl. 1983;17:105‐114. [PubMed] [Google Scholar]
- 7. Palta JR, Yeung DK, Frouhar V. Dosimetric considerations for a multileaf collimator system. Med Phys. 1996;23(7):1219‐1224. 10.1118/1.597678 [DOI] [PubMed] [Google Scholar]
- 8. Das IJ, Desobry GE, McNeeley SW, Cheng EC, Schultheiss TE. Beam characteristics of a retrofitted double‐focused multileaf collimator. Med Phys. 1998;25(9):1676‐1684. 10.1118/1.598348 [DOI] [PubMed] [Google Scholar]
- 9. Boyer AL, Ochran TG, Nyerick CE, Waldron TJ, Huntzinger CJ. Clinical dosimetry for implementation of a multileaf collimator. Med Phys. 1992;19(5):1255‐1261. 10.1118/1.596757 [DOI] [PubMed] [Google Scholar]
- 10. Klein EE, Harms WB, Low DA, Willcut V, Purdy JA. Clinical implementation of a commercial multileaf collimator: dosimetry, networking, simulation, and quality assurance. Int J Radiat Oncol Biol Phys. 1995;33(5):1195‐1208. 10.1016/0360-3016(95)00198-0 [DOI] [PubMed] [Google Scholar]
- 11. Vadash P, Bjärngard B. An equivalent‐square formula for head‐scatter factors. Med Phys. 1993;20(3):733‐734. 10.1118/1.597024 [DOI] [PubMed] [Google Scholar]
- 12. Kim S, Zhu TC, Palta JR. An equivalent square field formula for determining head scatter factors of rectangular fields. Med Phys. 1997;24(11):1770‐1774. 10.1118/1.597963 [DOI] [PubMed] [Google Scholar]
- 13. Kennedy C, Freedman G, Taunk N, et al. Whole breast irradiation with Halcyon™ 2.0: workflow and efficiency of field‐in‐field treatment with dynamic beam flattening technique and kV cone beam computed tomography. Cureus. 2018;10(10):e3510. 10.7759/cureus.3510 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Tamura M, Matsumoto K, Otsuka M, Monzen H. Plan complexity quantification of dual‐layer multi‐leaf collimator for volumetric modulated arc therapy with Halcyon linac. Phys Eng Sci Med. 2020;43(3):947‐957. 10.1007/s13246-020-00891-2 [DOI] [PubMed] [Google Scholar]
- 15. Choi MG, Law M, Yoon DK, et al. Simplified sigmoidal curve fitting for a 6 MV FFF photon beam of the Halcyon to determine the field size for beam commissioning and quality assurance. Radiat Oncol. 2020;15(1):273. 10.1186/s13014-020-01709-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Bollinger D, Laugeman E, Li T, et al. Technical Note: Dosimetric characterization of the dynamic beam flattening MLC sequence on a ring shaped, Jawless Linear Accelerator with double stacked MLC. Med Phys. 2020;47(3):948‐957. 10.1002/mp.14001 [DOI] [PubMed] [Google Scholar]
- 17. Laugeman E, Heermann A, Hilliard J, et al. Comprehensive validation of Halcyon 2.0 plans and the implementation of patient specific QA with multiple detector platforms. J Appl Clin Med Phys. 2020;21(7):39‐48. 10.1002/acm2.12881 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Lim TY, Dragojević I, Hoffman D, Flores‐Martinez E, Kim GY. Characterization of the Halcyon™ multileaf collimator system. J Appl Clin Med Phys. 2019;20(4):106‐114. 10.1002/acm2.12568 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Kim MM, Bollinger D, Kennedy C, et al. Dosimetric characterization of the dual layer MLC system for an O‐ring linear accelerator. Technol Cancer Res Treat. 2019;18:1533033819883641. 10.1177/1533033819883641 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Varian Medical Systems . Eclipse 16.1 Photon and Electron Reference Guide. P1044594‐001‐A. Varian Medical Systems; 2020. [Google Scholar]
- 21. Thatcher M, Bjärngard BE. Head‐scatter factors in rectangular photon fields. Med Phys. 1993;20(1):205‐206. [DOI] [PubMed] [Google Scholar]
- 22. Sterling TD, Perry H, Katz L. Derivation of a mathematical expression for the percent depth dose surface of cobalt 60 beams and visualization of multiple field dose distributions. Br J Radiol. 1964;37:544‐550. 10.1259/0007-1285-37-439-544 [DOI] [PubMed] [Google Scholar]
- 23. Boyer A, Biggs P, Galvin J, et al. Basic Applications of Multileaf Collimators: Report of the Task Group No. 50. Radiation Therapy Committee AAPM Report 72. AAPM; 2001. [Google Scholar]
- 24. Nara T, Iwasaki A, Komai F, Sato K, Fukushi H, Horanai Y. Two approximations to obtain the collimator scatter factor (S c) for MLC irregular fields. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2006;62(7):961‐970. 10.6009/jjrt.62.961 [DOI] [PubMed] [Google Scholar]
- 25. Onai Y, Irifune T, Tomaru T, Uchida I, Matsuda T. Applicability of area/perimeter method for determining tissue‐air ratios to irregular fields defined by multi‐split collimator for conformation therapy. Jpn Radiol Phys. 1981;1(1):11‐16. 10.11323/jjmp1981.1.1_11 [DOI] [Google Scholar]
- 26. Clarkson JR. A note on depth doses in fields of irregular shape. Br J Radiol. 1941;14:265‐268. 10.1259/0007-1285-14-164-265 [DOI] [Google Scholar]
- 27. Gay SS, Netherton TJ, Cardenas CE, et al. Dosimetric impact and detectability of multi‐leaf collimator positioning errors on Varian Halcyon. J Appl Clin Med Phys. 2019;20(8):47‐55. 10.1002/acm2.12677 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Richter A, Wegener S, Breuer K, et al. Comparison of sliding window and field‐in‐field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems. Radiat Oncol. 2021;16:213. 10.1186/s13014-021-01942-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Miften M, Olch A, Mihailidis D, et al. Tolerance limits and methodologies for IMRT measurement‐based verification QA: recommendations of AAPM Task Group No. 218. Med Phys. 2018;45(4):e53‐e83. 10.1002/mp.12810 [DOI] [PubMed] [Google Scholar]
- 30. Takahashi R, Kamima T, Itano M, et al. A multi‐institutional study of secondary check of treatment planning using Clarkson‐based dose calculation for three‐dimensional radiotherapy. Phys Med. 2018;49:19‐27. 10.1016/j.ejmp.2018.04.394 [DOI] [PubMed] [Google Scholar]
- 31. Georg D, Nyholm T, Olofsson J, et al. Clinical evaluation of monitor unit software and the application of action levels. Radiother Oncol. 2007;85(2):306‐315. 10.1016/j.radonc.2007.04.035 [DOI] [PubMed] [Google Scholar]
- 32. Hrbacek J, Lang S, Klöck S. Commissioning of photon beams of a flattening filter‐free linear accelerator and the accuracy of beam modeling using an anisotropic analytical algorithm. Int J Radiat Oncol Biol Phys. 2011;80(4):1228‐1237. 10.1016/j.ijrobp.2010.09.050 [DOI] [PubMed] [Google Scholar]