Skip to main content
. 2023 Jan 10;10:1090281. doi: 10.3389/fbioe.2022.1090281

TABLE 2.

Comparison of the performance of some of the LFIC test strips authorized for nucleic acid (2022c) by the U.S. Food and Drug Administration emergency use.

Product name Technology Target Characteristic PPV,% (95% CI) NPV,% (95% CI) Detection limit, copies/μL Testing time Reference
ID NOW COVID-19 Isothermal nicking enzyme amplification reaction RdRp gene Qualitative 94.5 (85.1–98.1) 99.3 (96.4–99.9) .511 5–13 min Babic et al. (2021)
Sherlock CRISPR SARS-CoV-2 kit RT-LAMP,CRISPR-Cas13a system N and ORF1ab gene Quantitative 100% accuracy compared to RT-PCR 6.75 50 min Khan et al. (2021)
SARS-CoV-2 RNA DETECTR Assay RT-LAMP,CRISPR-Cas12 system N and E gene Qualitative 95 (/) 100 (/) 10 45 min Broughton et al. (2020)
Cue COVID-19 test Isothermal nucleic acid amplification tests technology N gene Qualitative 91.7 (/) 98.4 (/) After 20 min Donato et al. (2021)