Table 2.

Comparison of sensitivity and specificity of stool antigen detection kits for amoebiasis

kit Intestinal amoebiasis
Sensitivity Specificity
Manufacturer Compared to Ref
TechLab Kit II 86-95% 93-100% TechLab, Blacksburg, VA, USA Microscopy and culture 20, 38, 105
71-79% 96-100% TechLab, Blacksburg, VA, USA Real-time PCR 20, 37, 106, 107
87.5% 100% TechLab, Blacksburg, VA, USA isoenzyme analysis
14.2% 98.3% TechLab, Blacksburg, VA, USA Culture and Isoenzyme
Entamoeba CELISA-PATH 95-100% 93-100% Cellabs Pty Ltd., Brookvale, Australia isoenzyme analysis
28 100 Cellabs Pty Ltd., Brookvale, Australia PCR
27.8 98 Cellabs Pty Ltd., Brookvale, Australia microscopy
Optimum S kit 100 Unknown Merlin Diagnostika, Berheim-Hersel, Germany ProSpecT Entamoeba histolytica microplate assay
ProSpecT Entamoeba histolytica
microplate assay; Alexon-Trend, Ramsey, MN
54.5 94 REMEL Inc., Lenexa, KS Culture and Isoenzyme 39,110
78% 99% REMEL Inc., Lenexa, KS microscopy
Triage parasite panel 100% 100% BIOSITE Diagnostics, San Diego, CA microscopy
96% 99.1% BIOSITE Diagnostics, San Diego, CA Permanent
staining
68.3% 100% BIOSITE Diagnostics, San Diego, CA ProSpecT Entamoeba histolytica microplate assay
RIDA®QUICK
Cryptosporidium/Giardia/Entamoeba Combi
100% 88% R-BioPharm, Darmstadt, Germany microscopy
62.5% 96.1% R-BioPharm, Darmstadt, Germany PCR
RIDA®QUICK
Entamoeba
28.6% 86.1% R-BioPharm, Darmstadt, Germany Techlab E. histolytica II
ImmunoCard STAT!® CGE 88% 92% Meridian Bioscence, Milan, Italy real-time PCR
E. HISTOLYTICA QUIK CHEK 78.4% 97% TechLab, Blacksburg, VA, USA ProSpecT Entamoeba histolytica microplate assay
98%- 100% 100% BIOSITE Diagnostics, San Diego, CA Techlab E. histolytica II 49, 50
44.7% 99.8% BIOSITE Diagnostics, San Diego, CA PCR