Table 2.
kit | Intestinal amoebiasis Sensitivity Specificity |
Manufacturer | Compared to | Ref | |
---|---|---|---|---|---|
TechLab Kit II | 86-95% | 93-100% | TechLab, Blacksburg, VA, USA | Microscopy and culture | 20, 38, 105 |
71-79% | 96-100% | TechLab, Blacksburg, VA, USA | Real-time PCR | 20, 37, 106, 107 | |
87.5% | 100% | TechLab, Blacksburg, VA, USA | isoenzyme analysis | ||
14.2% | 98.3% | TechLab, Blacksburg, VA, USA | Culture and Isoenzyme | ||
Entamoeba CELISA-PATH | 95-100% | 93-100% | Cellabs Pty Ltd., Brookvale, Australia | isoenzyme analysis | |
28 | 100 | Cellabs Pty Ltd., Brookvale, Australia | PCR | ||
27.8 | 98 | Cellabs Pty Ltd., Brookvale, Australia | microscopy | ||
Optimum S kit | 100 | Unknown | Merlin Diagnostika, Berheim-Hersel, Germany | ProSpecT Entamoeba histolytica microplate assay | |
ProSpecT Entamoeba histolytica microplate assay; Alexon-Trend, Ramsey, MN |
54.5 | 94 | REMEL Inc., Lenexa, KS | Culture and Isoenzyme | 39,110 |
78% | 99% | REMEL Inc., Lenexa, KS | microscopy | ||
Triage parasite panel | 100% | 100% | BIOSITE Diagnostics, San Diego, CA | microscopy | |
96% | 99.1% | BIOSITE Diagnostics, San Diego, CA | Permanent staining |
||
68.3% | 100% | BIOSITE Diagnostics, San Diego, CA | ProSpecT Entamoeba histolytica microplate assay | ||
RIDA®QUICK Cryptosporidium/Giardia/Entamoeba Combi |
100% | 88% | R-BioPharm, Darmstadt, Germany | microscopy | |
62.5% | 96.1% | R-BioPharm, Darmstadt, Germany | PCR | ||
RIDA®QUICK Entamoeba |
28.6% | 86.1% | R-BioPharm, Darmstadt, Germany | Techlab E. histolytica II | |
ImmunoCard STAT!® CGE | 88% | 92% | Meridian Bioscence, Milan, Italy | real-time PCR | |
E. HISTOLYTICA QUIK CHEK™ | 78.4% | 97% | TechLab, Blacksburg, VA, USA | ProSpecT Entamoeba histolytica microplate assay | |
98%- 100% | 100% | BIOSITE Diagnostics, San Diego, CA | Techlab E. histolytica II | 49, 50 | |
44.7% | 99.8% | BIOSITE Diagnostics, San Diego, CA | PCR |