Table 3.
Mean (SD) values for breathing rate and concurrent validity measured with the Breath Counter app and stethoscope methods.
| Stethoscope method, median (95% CI) | App method, median (95% CI) | Intraclass correlation coefficient (95% CI) | P valuea | Pearson r (P value) | R² | Limits of agreement (95% CI) | Coefficient of variation, % | Coefficients of variation of method errors | Repeatability coefficient | Repeatability coefficient, % |
| 15 (15.00 to 15.66) | 14.66 (14.25 to 15.07) |
0.793 (0.59 to 0.887) | .08 | 0.722 (<.001) |
0.521 (<.001) | −0.5 (−2.80 to 1.70) | 2.58 | 1.83 | 0.48 | 1.90 |
aA Mann-Whitney U test for independent samples was performed. In all analyses, P<.05 (with a 95% CI) was considered significant.