Skip to main content
. 2023 Jan 27;22:33. doi: 10.1186/s12936-023-04446-0

Table 2.

Malaria Screener and PVF-Net evaluation using microscopy as a reference

Evaluation Metrics Accuracy, % (95% CI) Sensitivity, % (95% CI) Specificity, % (95% CI)
Index tests
 Malaria Screener (85 patients) 74.1 (63.5—83.0) 100 (91.2—100) 51.1 (35.8—66.3)
 PVF-Net (189 patients) 83.1 (77.0—88.1) 86.9 (78.6—92.8) 78.9 (69.0—86.8)