Table 2.

MRI acquisition parameters for each center participating in the INNI initiative

Center A Group B Group C Group D
Scanner Philips Intera GE Signa HDxt Siemens Verio Philips Achieva
Coil 8-channel coil 8-channel coil 12-channel coil 32-channel coil
3D-T1 weighted
 Sequence [FFE–TFE] IR-FSPGR MPRAGE TFE
 Imaging plane [Axial–Sagittal] Sagittal Sagittal Axial
 Spatial resolution [mm3] [0.9 × 0.9 × 1.6–1 × 1 × 1] 1 × 1 × 1.2 0.5 × 0.5 × 1 1 × 1 × 2
 Acquisition matrix [256 × 256–256 × 256] 256 × 256 256 × 256 256 × 256
 Slices [220–204] 166 176 192
 TR [ms] [25–7] 6.9 1900 10
 TE [ms] [4.6–3.1] 2.8 2.9 4
 TI [ms] 650 900
 Flip angle [°] [30–9] 8 [8–9] 8
RS-fMRI
 Sequence EPI EPI EPI EPI
 Imaging plane Axial Axial Axial Axial
 Spatial resolution [mm3] 1.875 × 1.875 × 4 4 × 4 × 4 3 × 3 × 3 1.875 × 1.875 × 4
 Acquisition matrix 128 × 128 64 × 64 64 × 64 128 × 128
 Slices 30 29 50 30
 TR [ms] 3000 1500 3000 3000
 TE [ms] 35 32 30 35
 Flip angle [°] 90 90 89 90
 Volumes 200 240 140 200