Table 1.

Results of the baseplate navigation accuracy using augmented reality through head-mounted display

ID 3D vector Entry point Attempts Side
1 1.0° 3.3 mm 1 Left
2 3.2° 5.7 mm 1 Left
3 7.8° 2.3 mm 2 Left
4 3.2° 4.3 mm 1 Right
5 3.4° 4.0 mm 1 Right
6 4.7° 0.6 mm 1 Left
7 4.3° 2.3 mm 1 Left
8 4.9° 5.4 mm 1 Left
9 4.7° 1.9 mm 1 Left
10 3.4° 4.0 mm 1 Right
11 1.0° 6.2 mm 1 Left
12 3.3° 1.6 mm 1 Right

Mean

95% CI

3.8° ± 1.7°

(95% CI 2.6°; 4.9°)

3.5 ± 1.7 mm

(95% CI 2.4; 4.6)

1 ± 0.3

The values 3D vector and entry point describe the deviation from the planning measured by comparing the preoperatively planned vector with the postoperatively achieved vector using computed tomography. Attempts reflect the number of registration processes to achieve adequate hologram overlay

CI confidence interval, mm millimeter