Rapid GC-MS as a Screening Tool for Forensic Fire Debris Analysis

Briana A Capistran a, Edward Sisco a
PMCID: PMC9888146  NIHMSID: NIHMS1858175  PMID: 36733494

Abstract

Techniques developed for the screening of forensic samples can be useful for increasing sample throughput and decreasing backlog in forensic laboratories. One such technique, rapid gas chromatography mass spectrometry (GC-MS), allows for fast sample screening (≈1 min) and has gained interest in recent years for forensic applications. This work focuses on the development of a method for ignitable liquid analysis using rapid GC-MS. A sampling protocol and temperature program were developed for the analysis of these volatile samples. Using the optimized method for analysis, the limits of detection for compounds commonly found in ignitable liquids ranged from 0.012 mg/mL to 0.018 mg/mL. Once the method was developed, neat ignitable liquids (i.e., gasoline and diesel fuel) were analyzed, and major components in each liquid were identified. The identification of major compounds in gasoline and diesel fuel in the presence of substrate interferences was then assessed through the analysis of simulated fire debris samples. Three different substrates were spiked with each ignitable liquid, burned, and analyzed. Major compounds in both liquids were identified using the total ion chromatograms, relevant extracted ion profiles, and deconvolution methods.

Keywords: GC-MS, fire debris analysis, ignitable liquid residue, screening methods

1. Introduction

Ignitable liquid residue analysis is one of the most commonly requested trace evidence analyses in forensic laboratories [1, 2]. These liquids are often used to increase the speed and spread of fires, and the presence of ignitable liquid residues in fire debris samples can be useful when determining if a fire was intentionally set. Gas chromatography mass spectrometry (GC-MS) is the standard technique used for fire debris analysis due to its ability to separate and identify components in complex sample matrices [2, 3]. Traditionally, prior to instrumental analysis, samples are prepared using an approved extraction technique, such as passive-headspace extraction [4]. Upon GC-MS analysis, the chromatographic profiles are visually compared to reference chromatograms of common ignitable liquids. While the presence of substrate interferences can hinder identification of compounds found in ignitable liquids, the use of extracted ion profiles (EIPs) can minimize these interferences and aid in the comparison process.

Though effective, GC-MS analysis can be time-consuming in terms of sample pretreatment and instrumental analysis. First, sample preparation methods such as passive-headspace extraction often require several hours (up to 24 hours) of preparation time to desorb unknown volatiles [4]. While analysts often remedy this by heating samples in ovens overnight, this portion of the analysis workflow is still time intensive. Second, typical GC-MS methods for fire debris analysis are lengthy (e.g., 30 min). Depending on how many samples are submitted and if replicate analyses are performed, sequences can run for tens of hours or more for a single case. In total, this analysis process is costly in terms of time and resources, especially if not every sample contains ignitable liquid residue(s). Thus, to decrease total analysis times, it is necessary to reduce the time required for both sample preparation and instrumental analysis. To initiate this process, recent efforts have focused on the development and application of screening techniques for fire debris applications to decrease instrumental analysis times.

Various analytical methods have been developed for screening and identification of ignitable liquids, many of which employ standalone spectrometric methods and eliminate the use of separation techniques. Such methodologies include headspace-mass spectrometry (HS-MS) [5–7], ion mobility spectrometry (IMS) [8, 9], and direct analysis in real-time mass spectrometry (DART-MS) [10]. Due to the volatilities of components in fire debris samples, HS-MS has been demonstrated as a useful technique for fast desorption and analysis of such components. Recent developments involve the application of an HS-MS method combined with chemometric tools for ignitable liquid identifications and classification [5–7]. Neat liquids and simulated fire debris samples were heated to desorb volatile sample components, and the sample headspace was collected via syringe and directly injected into a mass spectrometer. Using multivariate statistics, samples containing ignitable liquids were differentiated from those without ignitable liquids. The liquids in many of the positive samples were also correctly classified based on the ignitable liquid classes defined in ASTM E1618 (e.g., isoparaffinic, petroleum distillate, aromatic) [3], which are used for ignitable liquid identifications. While this technique enables rapid sample analysis, the lack of a separation component results in complicated spectra, as one spectrum is representative of all sample components. Due to the complexity of fire debris samples, the intricacies of resulting spectra can vary drastically and complicate the identification process.

Rapid analysis of ignitable liquids has also been studied using IMS and DART-MS, both of which contain atmospheric pressure components (ionization or mass analysis) and enable sample screening in real time. Similar to the HS-MS methods, IMS detection has been used for sample aliquots taken from the headspace of heated fire debris samples [8, 9]. In one such study, IMS sum spectra (IMSSS) were generated for samples containing ignitable liquids of four different ASTM classes. Chemometric tools were applied to the IMSSS data to classify samples based on drift time and intensity. Spectral fingerprints for each of the ignitable liquids were determined using additional multivariate statistical methods. The presence and class of ignitable liquids in 80 simulated fire debris samples were determined using these fingerprints as classification parameters. Despite the success of this work, not all forensic laboratories have the instrumentation necessary for performing analyses according to this method.

DART-MS, however, has gained popularity in forensic laboratories as a screening tool for samples such as seized drugs and explosives [11]. The increasing use of this technique coupled with thermal desorption (TD) for forensic sample analysis has prompted the application to fire debris samples. In this coupled method, temperature-programmed TD acts as a pseudo separation technique without requiring the use of lengthy chromatographic methods. While still complex, the resulting spectra are less so than those of samples without programmed desorption [12]. In one such study, TD-DART-MS was used for the desorption of volatile sample components in neat ignitable liquids and simulated fire debris samples [10]. Characteristic ions were identified for compounds in both gasoline and diesel fuel, and those same ions were used for identification of the liquids’ components in simulated debris samples. Due to the high temperatures reached in the TD temperature program (≈600 °C), TD-DART-MS was beneficial for identifying nonvolatile gasoline additives that could be used as differentiating ions in the presence of substrate interferences [10]. This technique enables rapid screening of ignitable liquids with minimal to no sample preparation; however, the resulting spectra, while less complex than those from other spectrometric techniques, can still be complicated to interpret, especially for identification of ignitable liquids that have evaporated to varying extents.

In addition to spectrometric methods, rapid chromatographic techniques have gained popularity in recent years as a screening tool for various samples, especially for forensic applications. Such techniques enable sample separation using a short chromatography column (2 m), followed by detection using conventional electron ionization mass spectrometry [13]. While rapid GC-MS methods are not designed to provide complete baseline separation due to the timescale of the technique (≈1 minute), separation is sufficient for preliminary sample screening [13]. In this work, the utility of rapid GC-MS as a screening tool for fire debris analysis was investigated. Prior to this work, the benchtop-based technique had been successfully demonstrated for the analysis of seized drugs and explosives [13]; however, extensive application to ignitable liquid residue analysis has not yet been performed. As such, focus in the first portion of this study was placed on developing a sampling procedure and temperature program specifically for ignitable liquids and volatile compounds. Method validation was performed, and limits of detection were determined for compounds commonly present in ignitable liquids. Finally, neat ignitable liquids (i.e., gasoline and diesel fuel) and simulated fire debris samples were analyzed. As with the other rapid techniques described above, this work focuses on the second component of fire debris analysis (instrumental analysis) in an effort to decrease analysis times and increase sample throughput. In order to fully achieve a rapid analysis, additional studies focused on different sample preparation approaches would need to be completed.

2. Materials and Methods

2.1. Rapid GC-MS

All analyses were completed using an Agilent 3971 QuickProbe GC-MS system coupled to an Agilent 8890 gas chromatograph and a 5977B mass spectrometer, equipped with a 7693 autosampler (Agilent Technologies, Santa Clara, CA, USA). Configuration of the rapid GC-MS column such that the MS could be used by both the rapid and traditional liquid injection GC was done using an inert splitter (Supplementary Data, Figure S1). The bottom port of the splitter was connected to the MS through a DB-1ms Ultra Inert QuickProbe GC column (1 m length × 0.18 mm outer diameter × 0.18 μm inner diameter). The middle port was connected to the rapid GC column, a DB-1ht QuickProbe GC column (2 m length × 0.25 mm outer diameter × 0.10 μm inner diameter), which is housed inside the rapid system attachment. The top port of the splitter was connected to the traditional, liquid-injection GC column, a 100 % polydimethylsiloxane column (DB-1, 30 m length × 0.25 mm outer diameter × 0.25 μm inner diameter, Agilent Technologies). For both rapid and traditional GC, helium (99.999 %) was used as the carrier gas with a flow rate of 1 mL/min.

2.2. Sampling protocol and instrumental analysis

A seven-component standard mixture used for method development consisted of the following compounds: p-xylene (≥99 %, Sigma-Aldrich, St. Louis, MO, USA), n-nonane (C9, >98.0 %, TCI Chemicals, Tokyo, Japan), 1,2,4-trimethylbenzene (98 %, Sigma-Aldrich), n-decane (C10, ≥99 %, Sigma-Aldrich), 1,2,4,5-tetramethylbenzene (98 %, Sigma-Aldrich), 2-methylnaphthalene (analytical standard, Sigma-Aldrich), and n-tridecane (analytical standard, Sigma-Aldrich). All compounds were prepared at a concentration of 1 mg/mL in dichloromethane (99.9 %, Sigma-Aldrich).

For rapid GC-MS analysis, the optimized temperature program and analysis method is described in detail in Section 3.1. The instrument parameters used are summarized in Table 1. A solvent delay was not used, as analysis timescales are faster than detector on/off response times. The GC oven was held isothermal at 280 °C to prevent recondensation of the analyte in the transfer column from the splitter to the MSD. For all rapid GC-MS samples, the order of analyses was as follows: system blank (no sample or probe inserted), probe (or solvent) blank, sample(s), probe (or solvent) blank, system blank.

Table 1.

Parameters for optimized rapid GC-MS temperature program and analysis method.

Temperature program 35 °C, hold 10 s
2 °C/s to 100 °C
4 °C/s to 280 °C, hold 10 s
Total run time 98 s (1.63 min)
Inlet temperature 250 °C
Injection mode Split (10:1 fixed)
Injection time 7 s
Liner type Ultra Inert Fritted Liner (Agilent Technologies)
Carrier gas He, 103.421 kPa (15 psi)
GC oven temperature 280 °C
Transfer line 280 °C
MSD source Electron ionization, 70 eV
Source temperature 230 °C
Quadrupole temperature 150 °C
Tune type atune
Scan range m/z 40 – m/z 550
Threshold 50 counts
Scan rate N = 1 (5.9 scans/s)
Gain factor 3 arbitrary units (a.u.)

To validate the rapid GC-MS results, all samples were also analyzed using traditional GC-MS. The injection port was set to 250 °C. For all samples, a 1 μL injection was used, with a split ratio of 20:1 and a pressure of 48.745 kPa (7.0699 psi). The temperature program was based on previous methods for fire debris analysis [14, 15]: initial temperature of 40 °C, hold for 3 min, ramp 10 °C/min to 280 °C, hold for 4 min (total run time of 31 min). A 5 min solvent delay was utilized. The remaining parameters for MSD detection were the same as those used for rapid GC-MS analysis. For traditional GC-MS samples, the order of analyses was as follows for each set of sample replicates: solvent blank, sample(s), solvent blank.

2.3. Data processing

All data were collected and analyzed using MassHunter (MassHunter Workstation Software, GC-MS Data Acquisition, version 10.0, Build 10.0.284.1; Qualitative Analysis Navigator, version B.08.00, Agilent Technologies). Compounds were identified using the NIST 17 Mass Spectral Library. Extracted ion profiles (EIPs) were generated as needed using the m/z values of the following major compound classes present in gasoline and/or diesel fuel: alkane (m/z 57, m/z 71, m/z 85, m/z 99), aromatic (m/z 91, m/z 105, m/z 119, m/z 133), cyclic alkane (m/z 55, m/z 69, m/z 83, m/z 97), indane (m/z 117, m/z 131, m/z 145, m/z 159), and polynuclear aromatic (m/z 128, m/z 142, m/z 156). Deconvolution was performed as needed using MassHunter Unknowns Analysis (MassHunter Workstation Software, Quantitative Analysis, Unknowns Analysis, version B.09.00/Build 9.0.647.0). Parameters used for deconvolution are listed in Supplementary Data, Table S1. Raw GC-MS (rapid and traditional) data files can be found on the NIST Public Data Repository: https://doi.org/10.18434/mds2-2601.

2.4. Limit of detection (LOD) determination

Limit of detection (LOD) measurements were made following the procedure outlined in ASTM E2677 [16] using n-decane as the analyte of interest. Five nominal concentrations of 1 mg/mL, 0.5 mg/mL, 0.1 mg/mL, 0.05 mg/mL, and 0.01 mg/mL were chosen, and solutions were prepared gravimetrically. Ten replicates of each concentration along with ten solvent blanks were analyzed by rapid and traditional GC-MS. The five concentrations were analyzed in random order determined using the random number generator in Microsoft Excel (Microsoft Office 365, Microsoft Corporation, Redmond, WA, USA). For each concentration, abundance values for the replicate measurements were taken at the peak apices and used to generate a gravimetric calibration curve. The LODs were calculated using the NIST ASTM E2677 Limit of Detection Web Portal Data Entry Page [16] and the default settings, which provide a 90 % confidence level.

2.5. Mock burn sample collection, preparation, and analysis

Mock burn samples were collected at the NIST Fire Department facilities. Pieces of carpet (medium pile, approximately 5 cm × 6 cm), unfinished wood (approximately 4 cm × 5 cm × 4 cm), and primed wood (approximately 4 cm × 7 cm × 2 cm) were used as substrates. Substrates were analyzed as either (i) unburnt, (ii) burnt, no ignitable liquid present, or (iii) burnt, ignitable liquid present. Gasoline and diesel fuel (purchased locally) were provided by the Fire Department and used as ignitable liquids.

Unburnt substrates (i) were analyzed without further preparation. For burnt, no ignitable liquid present substrates (ii), individual pieces of substrates were burned using a small amount (< 1 mL) of isopropyl alcohol for ignition. Because of the fire-resistant nature of the substrates, other ignition methods (e.g., torch) were not sufficient to maintain burning of the burnt, no ignitable liquid present substrates (ii). The high volatility of isopropyl alcohol enables it to be used for initial ignition, burning completely before flames self-extinguish and leaving the bare substrate to burn. Each substrate was allowed to burn until flames were no longer present. For burnt, ignitable liquid present substrates (iii), a 1 mL-aliquot of gasoline or diesel fuel was spread onto fresh pieces of each substrate, subsequently ignited, and allowed to burn until flames were no longer present. Replicates of each burn type and substrate type combination were conducted. All samples were packaged in separate, unlined, metal pint-sized paint cans (Uline, Pleasant Prairie, WI, USA).

Samples were prepared for instrumental analysis using the ASTM E1412-16 procedure for passive-headspace extraction [4]. Briefly, a third of an activated charcoal strip (Albrayco Technologies Inc., Cromwell, CT, USA) was suspended above the burn sample in each paint can, and the samples were heated at 80 °C for four hours. A temperature of 80 °C was chosen to maximize adsorption of volatiles onto the charcoal strips, especially those of higher molecular weight (compounds heavier than C16) [4]. After this time, the samples were removed from the oven and allowed to cool to room temperature. Each charcoal strip was removed and washed with 300 μL of dichloromethane and vortexed for 10 s. For instrument replicates, the eluent was divided among separate gas chromatography vials to minimize evaporation of volatile sample components. Duplicate replicates of all samples were then analyzed by rapid and traditional GC-MS. Unburnt substrates without ignitable liquids were also analyzed using the same procedure. In addition to the burn samples, the neat ignitable liquids were each prepared at an approximate concentration of 5 mg/mL and analyzed by rapid and traditional GC-MS.

3. Results and Discussion

3.1. Sampling protocol and rapid GC-MS method optimization

3.1.1. Probe tip versus direct injection

For sampling protocol determination and method development, a seven-component standard mixture was prepared consisting of compounds commonly found in ignitable liquids. Prior to rapid GC-MS analysis, the mixture was analyzed by traditional GC-MS to confirm the identities of all components and determine elution order. A chromatogram of the mixture analyzed using traditional GC-MS is shown in Supplementary Data, Figure S2.

As ignitable liquids are volatile mixtures prone to evaporation, the effectiveness of the probe holder fitted with the round tip probe was first investigated. For this portion of the study, a generic (unoptimized) temperature program was utilized (Supplementary Data, Table S2). The probe tip was dipped into the standard mixture solution and subsequently inserted into the system inlet. Solvent evaporation was observed on the probe tip in the time between sampling and injection. In the resulting chromatogram, dichloromethane, p-xylene, and n-nonane were present at low abundances compared to the last four eluting compounds (Figure 1A). The low abundances of p-xylene and n-nonane, the two most volatile analytes in the mixture, were attributed to evaporation with the solvent prior to sample introduction into the instrument. While solvent evaporation is recommended for nonvolatile samples such as seized drugs [13, 17], this evaporation will affect the profile of volatile samples such as ignitable liquids.

Figure 1.

Figure 1.

(A) Total ion chromatogram (TIC) of original probe tip injection of the standard mixture prior to sampling modifications, (B) TICs of replicate probe tip injections of the standard mixture, (C) TIC of the standard mixture using direct liquid injection, and (D) average peak heights for all seven compounds in the standard mixture across replicate injections using the probe tip (light blue) and direct injection (dark blue) methods (n = 4). Note that retention times past 0.5 min are not shown in (A) – (C) because no compounds eluted after this time. DCM = dichloromethane; TMB = trimethylbenzene; TEMB = tetramethylbenzene; MN = methylnaphthalene.

To prevent evaporation during sampling, the probe tip was then held in solution for approximately 10 s (to sample a larger volume), and sampling was performed directly adjacent to the instrument inlet to minimize time and evaporation between sampling and sample introduction. Greater abundances overall were observed with these modifications, but the abundances of the seven analytes in the mixture relative to each other did not appear accurate, as all compounds were prepared at the same concentration (Figure 1B). For example, average peak heights of p-xylene and n-nonane, the compounds with the lowest abundances, were approximately 40 % and 37 %, respectively, of the peak height of n-tridecane, one of the most abundant compounds. While some variation in peak heights among the seven compounds was expected due to differences in detector response factors and vapor pressures, peak heights less than 50 % of the most abundant peak were not expected. In addition to relative peak height, the reproducibility of abundances across multiple injections was poor using the probe tip (Figure 1B), with relative standard deviations (RSDs) ranging from ≈89 % (n-tridecane) to ≈130 % (p-xylene). Given the extent of coelution and lack of baseline resolution, peak height was used as the variable for comparison and RSD calculation, as they were extracted more reproducibly than peak areas. Based on these results, evaporation from the probe tip was determined to influence the chromatographic results for volatile samples.

Direct liquid injection was assessed as an alternative sampling technique. Using a 10 μL autosampler syringe, 1 μL of the standard mixture was injected directly into the inlet. Using this approach, the relative abundances between compounds was more reasonable than with the probe tip injection technique (Figure 1C). For example, the peak heights of p-xylene and n-nonane were ≈88 % and ≈72 %, respectively, of that of n-tridecane. A slight shift towards later retention times was observed despite use of the same temperature program as the probe tip method; this shift was attributed to the presence of a wider solvent peak due to the nature of direct injections. Comparatively, the abundances of the seven compounds using direct injection were approximately five to ten times greater than those using the probe tip method (Figure 1D). Peak heights were also more reproducible using the direct injection method, with RSDs ranging from ≈3 % (n-decane) to ≈8 % (n-tridecane). Based on these improvements, the direct injection method was utilized for the remainder of samples in this study.

3.1.2. Rapid GC-MS temperature program optimization and method development

Once an appropriate sampling method was established, a temperature program and analysis method were optimized for fire debris applications using the seven-component standard mixture. Parameters investigated included initial and final hold times, ramp rates, and temperatures. Gas flow was not optimized due to the fact that the column is under vacuum. Given that this technique is designed as a rapid analysis method intended for screening purposes, it was understood that baseline separation may not be possible and coelution may occur, especially for complex samples such as ignitable liquids. In this portion of the study, the goal was to optimize the separation between the components in the standard mixture as much as possible, while still maintaining rapid analysis times (less than two minutes). Focus was placed mainly on the first four compounds (i.e., p-xylene, n-nonane, 1,2,4-trimethylbenzene, n-decane), for which the greatest extent of coelution was observed among the seven components (Figure 1C) using the generic method.

The initial temperature program was modeled after the traditional GC program detailed in Section 2.1 using the following parameters: initial temperature of 40 °C, hold 7 s, ramp 7 °C/s to 280 °C, hold 15 s. Modifications were made to this original method to optimize separation, and the most effective changes are summarized here. Increasing the initial hold time to 10 s and decreasing the ramp rate to 4 °C/s improved separation and the extent of coelution (Supplementary Data, Figure S3, top). Additionally, as final hold time did not affect separation, it was decreased to 10 s to decrease overall run time for subsequent analyses.

Despite the improved separation, ignitable liquids often contain a greater number of compounds than in the standard mixture. Given that many of these compounds have similar volatilities and thus would elute in similar regions of a chromatogram, the temperature program was modified further to achieve increased separation. The effects of a second ramp on separation were explored to improve separation in the early portions of the chromatogram (retention time (tR) < 0.3 min). Following the initial 10 s hold, a ramp of 2 °C/s to 100 °C was added specifically for separation of highly volatile compounds found in ignitable liquids. The second ramp remained at 4 °C/s to 280 °C. Separation improved for the first four eluting compounds (Supplementary Data, Figure S3, bottom). The addition of the slower ramp (2 °C/s) also improved the separation of the last three eluting compounds, where baseline separation was achieved (Supplementary Data, Figure S3, bottom).

Because the solvent will always be present in a chromatogram for a direct injection sample, it is possible that highly volatile compounds could coelute with the solvent. To mitigate coelution as much as possible for compounds more volatile than those in the standard mixture, the initial temperature was decreased to 35 °C. While subtle, a slight improvement in separation was observed for p-xylene and n-nonane (Figure 2). This last temperature program was determined to be reproducible, with retention time RSDs among replicate injections ranging from 0.5 % to 2.1 % across the seven compounds. Based on the demonstrated reproducibility and improved separation achieved across all method modifications made, this temperature program was determined to be most optimal for ignitable liquids. The final temperature program and method parameters are summarized in Table 1 and were used for all subsequent analyses.

Figure 2.

Figure 2.

Total ion chromatogram (TIC) of the standard mixture analyzed using the optimized temperature program. DCM = dichloromethane; TMB = trimethylbenzene; TEMB = tetramethylbenzene; MN = methylnaphthalene.

3.2. Limit of detection (LOD) measurement

To effectively utilize a technique for screening purposes, the associated limit of detection (LOD) must be comparable, if not lower, than that of a confirmatory technique. As such, the LODs for both rapid and traditional GC-MS were measured using the method outlined in ASTM E2677 [16]. N-decane was used as the analyte of interest, as it is present in many ignitable liquids such as gasoline and diesel fuel and is neither the most volatile nor least volatile component. Five solutions of n-decane were prepared gravimetrically with calculated concentrations of 0.96 mg/mL, 0.48 mg/mL, 0.090 mg/mL, 0.047 mg/mL, and 0.0094 mg/mL (nominal concentrations of 1 mg/mL, 0.5 mg/mL, 0.1 mg/mL, 0.05 mg/mL, and 0.01 mg/mL, respectively). Upon instrumental analysis, a gravimetric calibration curve was generated by plotting abundance versus concentration for rapid GC-MS (Supplementary Data, Figure S4A) and traditional GC (Supplementary Data, Figure S4B). Increased variability in abundance between instrument replicates was observed for rapid GC-MS, which was expected and was attributed to random error associated with manual injection.

The estimated LOD for each method was calculated at the 90 % confidence level. For rapid GC-MS, the LOD of n-decane was calculated to be 0.018 mg/mL with a 90 % upper confidence limit of 0.028 mg/mL. For traditional GC, the LOD was estimated to be 0.013 mg/mL with a 90 % upper confidence limit of 0.016 mg/mL. The LOD for rapid GC-MS had a greater 90 % upper confidence limit and therefore greater uncertainty, which is reasonable considering the inherent variability associated with manual injection as opposed to automatic injection used for traditional GC analysis. The higher limit is not considered to be a limiting factor in the application of rapid GC-MS.

Having successfully determined the LOD of n-decane as a single-component mixture, LODs of the remaining six compounds in the standard mixture were calculated to investigate instrument performance across the analysis method temperature range. Solutions of the standard mixture were prepared at the same nominal concentrations used for analysis of n-decane. Calibration curves were generated for each compound and for each analysis method (Supplementary Data, Figures S5 – S6). Given that n-decane was still present in the standard mixture, the associated LOD was also calculated in addition to those for the six remaining components and compared to that calculated for the single-component analysis. For rapid GC-MS and traditional GC-MS, the LODs of n-decane in the standard mixture were estimated to be 0.012 mg/mL (0.014 mg/mL 90 % UCL) and 0.014 mg/mL (0.017 mg/mL 90 % UCL), respectively. These values were within the range of those reported for the single component analysis of n-decane for each respective analysis method. While the single-component n-decane LOD for rapid GC-MS had a greater 90 % UCL compared to that for the mixture analysis, this difference was likely due to variability in the manual injection technique as well as inter-day instrument variability.

The LODs for the remaining six compounds were similar to those of n-decane (single-component and mixture analysis) for each respective analysis technique and within the 90% confidence limits (Supplementary Data, Table S3). For rapid GC-MS, LODs ranged from 0.012 mg/mL to 0.015 mg/mL; for traditional GC-MS, LODs ranged from 0.013 mg/mL to 0.015 mg/mL (Figure 3). It is important to note that the LODs are for individual compounds commonly present in ignitable liquids and do not represent LODs for ignitable liquids in general. As ignitable liquids are complex mixtures, coelution and peak overlap is much more likely for these samples than in the standard mixture. The increased noise for such liquids would result in increased LODs, and as such, the values reported here can be treated as benchmarks as opposed to actual representations of ignitable liquid LODs. However, for the purposes of this work, determination of detection limits for compounds across a range of volatilities was useful for validation of rapid GC-MS analysis.

Figure 3.

Figure 3.

Limits of detection (LOD) calculated for the standard mixture components for rapid GC-MS (light red, left) and traditional GC-MS (dark red, right) using ASTM E2677. Error bars represent the 90 % confidence intervals. *Note: The LOD for n-decane is reported here as the calculation from analysis of the standard mixture (not single-component analysis).

3.3. Application to ignitable liquids and mock burn samples

3.3.1. Rapid GC-MS analysis of ignitable liquids

Ignitable liquids were subject to rapid GC-MS analysis to assess the ability to identify major compounds present. As proof of concept, gasoline and diesel fuel were used, as those are the two most commonly used ignitable liquids [1, 2]. Samples were analyzed using the optimized method with validation by GC-MS (Supplementary Data, Figure S7). The total ion chromatogram (TIC) of gasoline contained several wide, split peaks, especially at early retention times (tR < 0.4 min) (Figure 4A). Despite the lack of baseline separation, compounds characteristic of gasoline were identified using simple peak picking (no further data processing). For example, several C2-alkylbenzenes (i.e., o/p-xylene, ethylbenzene) were identified between 0.15 min and 0.25 min, and C3-alkylbenzenes (e.g., 1,2,4-trimethylbenzene, 1,3,5-trimethylbenzene) were identified between 0.25 min and 0.37 min based on high match scores (≥ 80 a.u.). Low match scores (< 60 a.u.) prevented definitive identifications of compounds at longer retention times (tR > 0.5 min). Such scores were likely due to the lack of separation and peak definition at these times.

Figure 4.

Figure 4.

(A) Total ion chromatogram (TIC) of gasoline analyzed using rapid GC-MS (inset: maximized view of TIC from tR = 0.2 min to 1.0 min), as well as representative extracted ion profiles (EIPs) generated for the (B) aromatic and (C) polynuclear aromatic classes. DCM = dichloromethane; TMB = trimethylbenzene; TEMB = tetramethylbenzene; MN = methylnaphthalene.

Additional compounds were identified upon generating extracted ion profiles (EIPs). Several C4-alkylbenzenes (e.g., diethylbenzene, p-cymene, tetramethylbenzene) were identified in the aromatic EIP (Figure 4B), while naphthalene and 2-methylnaphthalene were identified in the polynuclear aromatic EIP (Figure 4C). The alkane, indane, and cyclic alkane EIPs were also generated but only contained compounds of low abundances (< 102 counts) that were unable to be definitively identified. Deconvolution of the gasoline TIC was then performed using MassHunter Unknowns Analysis (see Supplementary Data, Table S1 for parameters) and enabled identification of additional characteristic components. Toluene, one of the major compounds in gasoline, was identified at tR = 0.13 min in the deconvoluted chromatogram (Supplementary Data, Figure S8); however, it coeluted with dichloromethane and was thus masked by the intensity of the solvent peak in the original TIC (prior to deconvolution). Branched alkanes, indane, naphthalene, and methylnaphthalene isomers, all characteristic compounds in gasoline, were also identified in the deconvoluted chromatogram (Supplementary Data, Figure S8). Based on these results, utilization of EIPs and deconvolution as data processing tools enabled identification of the major components found in gasoline.

The major compounds in diesel fuel were also identified upon rapid GC-MS analysis. In the TIC of diesel fuel, the n-alkanes present resembled a Gaussian profile (Figure 5), which is characteristic of diesel fuel and petroleum distillates [2]. At approximately tR = 0.75 min, a slight deviation in the repeating pattern of n-alkanes was observed, and the spacing between n-alkanes became shorter at tR > 0.75 min. This change in the chromatographic profile was attributed to the change in ramp rates in the temperature program, which began around this time. In addition to the TIC, the alkane EIP was useful for confirming the presence of n-alkanes, as noise was decreased (Supplementary Data, Figure S9). Branched alkanes eluting between the n-alkanes were also identified using the alkane EIP.

Figure 5.

Figure 5.

Total ion chromatogram (TIC) of diesel fuel analyzed using rapid GC-MS. Note that peaks labeled Cn denote saturated, straight-chain alkanes consisting of n carbons.

3.3.2. Identification of major ignitable liquid components in simulated fire debris samples

Simulated fire debris samples were then prepared and analyzed using rapid GC-MS. Carpet, unfinished wood, and primed wood were used as substrates, as those are commonly used building materials. For each sample, an aliquot of ignitable liquid (i.e., gasoline or diesel fuel) was spiked onto the substrate and subsequently ignited. All samples were allowed to burn until completion (until flames self-extinguished). Following sample preparation by passive-headspace extraction, the compound identities in the resulting chromatograms were determined to examine the possibility of identifying major compounds in each ignitable liquid.

The gasoline-spiked carpet burn sample was one of the samples that experienced the greatest degree of burning. In the corresponding TIC, volatile compounds were still present at early retention times (Figure 6A). The C2-, C3-, and C4-alkylbenzenes that were identified in neat gasoline (Figure 4) were identified in the burn sample between tR = 0.15 min and 0.40 min. The extent of coelution at tR > 0.40 min hindered compound identification using simple peak picking alone; however, using representative EIPs and deconvolution, additional compounds were identified. The presence of the C2-, C3-, and C4-alkylbenzenes were further confirmed in the aromatic EIP (Figure 6B). Indane and branched indenes were identified using the indane EIP (Figure 6C), and naphthalene and 2-methylnaphthalene were identified in the polynuclear aromatic EIP (Figure 6D). As EIPs are commonly used to minimize substrate interferences, the use of these profiles for this sample confirms their utility for this screening technique.

Figure 6.

Figure 6.

(A) Total ion chromatogram (TIC) of simulated burn carpet sample spiked with gasoline, as well as representative extracted ion profiles (EIPs) for the (B) aromatic, (C) indane, and (D) polynuclear aromatic classes. MN = methylnaphthalene. Note: all chromatograms are shown across the maximized range tR = 0.1 min to 1.2 min.

Applying the deconvolution tool to the gasoline-spiked carpet TIC enabled further compound identification. Toluene (tR = 0.13 min), indane (tR = 0.31 min), and branched alkanes (tR = 0.12 min, 0.34 min, 0.66 min, 0.73 min) were identified. Styrene and benzaldehyde (tR = 0.21 min and 0.27 min, respectively), which were present in the samples of unburnt and burnt carpet without gasoline (Supplementary Data, Figures S10A and S10B), were also identified in the simulated burn sample following deconvolution (Supplementary Data, Figure S10C). These substrate compounds were present at low abundances and eluted at similar retention times as other, more abundant compounds in the burn sample and thus were not identified using the original TIC alone. Despite the high degree of burning of this sample, many compounds present in neat gasoline (Figure 4) were successfully identified in the burn sample using the TIC and data processing tools. For the additional samples spiked with gasoline, major compounds present in the ignitable liquid were identified in the simulated burn samples using the unfinished and primed wood substrates (Supplementary Data, Figures S11 – S12). For each burn sample, the major compounds identified were characteristic gasoline components (e.g., alkylbenzenes, indane, naphthalene, methylnaphthalene isomers).

Identification of compounds spanning a wider carbon range was investigated through rapid GC-MS analysis of simulated burn samples containing diesel fuel. Of the three substrates used, the extent of burning was greatest for the carpet sample when spiked with diesel, similar to the burn samples with gasoline. Despite the extent of burning, a Gaussian pattern of n-alkanes was observed in the TIC of the carpet burn sample (Figure 7), which is a pattern characteristic of diesel fuel and other petroleum distillates. This pattern was more pronounced in the alkane EIP due to the decrease in noise (Figure 7, inset), which was used to further confirm the presence of n-alkanes. In the neat diesel fuel sample, n-alkanes as heavy as n-nonadecane (C19) were present (Figure 5), whereas in the burn sample, n-heptadecane (C17) was the heaviest n-alkane definitively identified. Despite the use of the same diesel fuel for the burn sample as in the neat liquid analysis, it is likely that these compounds did not fully desorb from the burn sample during passive-headspace extraction due to their lower volatilities. This hypothesis was confirmed through traditional GC-MS analysis of the burn sample, in which n-heptadecane (C17) was also the heaviest n-alkane identified (Supplementary Data, Figure S13).

Figure 7.

Figure 7.

Total ion chromatogram (TIC) of simulated burn carpet sample spiked with diesel fuel (inset: representative extracted ion profile (EIP) for the alkane class). DCM = dichloromethane. Note that peaks labeled Cn denote saturated, straight-chain alkanes consisting of n carbons.

Even though the major components in diesel fuel were identified using the TIC and alkane EIP, the application of deconvolution methods enabled additional identifications, specifically for coeluting compounds present in lower abundances. Aromatic compounds, such as alkylbenzenes, were identified between tR = 0.20 min and 0.34 min, and several branched alkanes were identified between the n-alkanes. In addition to the normally distributed n-alkanes, the presence of the lesser abundant compounds is important for the classification of diesel fuel as a petroleum distillate. Substrate-specific compounds were also identified through deconvolution of the burn sample TIC, such as styrene (tR = 0.17 min) and benzaldehyde (tR = 0.22 min). These compounds were present at low abundances and coeluted with the more abundant compounds present at similar retention times. Aside from the carpet burn sample, analyses of the unfinished and primed wood samples spiked with diesel shared similar results as the wood substrates spiked with gasoline, as major compounds in diesel fuel were identified in both samples (Supplementary Data, Figures S14 – S15).

Across all simulated burn samples collected and analyzed, characteristic compounds in each ignitable liquid were identified through the combination of the raw TIC, EIP(s), and/or deconvoluted TIC. For the gasoline-spiked samples, in which many of the compounds had similar volatilities and eluted over a narrow retention time window, the use of EIPs and deconvolution was instrumental for identifying coeluting compounds and compounds present at low abundances. While major compounds in both liquids were successfully identified in this work, it is worth noting that ignitable liquids in actual case samples may be present at much lower concentrations. In these instances, there exists the possibility that substrate interferences could hinder identification of ignitable liquid components, possibly leading to a false negative conclusion. While the use of EIPs and deconvolution methods can minimize this risk, additional work is necessary to analyze samples (simulated and real) that contain lower concentrations of ignitable liquids. However, for the purposes of this foundational study, ignitable liquid components in the debris samples were identified in the presence of varying degrees of substrate interferences using rapid GC-MS and a combination of data processing tools.

4. Conclusions

In this proof-of-concept study, ignitable liquids were analyzed using a rapid GC-MS approach. A sampling protocol and optimized temperature program were developed for direct liquid injection of these volatile samples. When applied to neat ignitable liquids, major compounds in both gasoline and diesel fuel were identified in the resulting TICs. Use of representative EIPs enabled identification of additional characteristic compounds. Deconvolution proved to be another useful tool for the identification of lesser abundant coeluting compounds. Application of rapid GC-MS to mock burn samples resulted in identification of major compounds in both ignitable liquids despite the presence of substrate components. While the optimized method was developed for the benchtop-based instrumental approach utilized in this work, there is the potential for the method and/or method concepts (e.g., two ramps, ramp rates) to be extended to field-portable GC-MS instrumentation that involve separation using a short (e.g., 2 m) column. Such applications are the focus of future work. Additionally, ongoing efforts to expand on this foundational study involve the analysis of liquids of different ASTM classes to more comprehensively gauge the robustness of the rapid GC-MS technique as well as studies focused on the analysis of low concentration samples. Application to these additional liquids will be useful to investigate the performance of this technique for liquid classifications. Using the optimized instrumental method, sample analysis times were under two minutes. Due to the use of liquid injection, sample preparation was still performed in a similar manner as is currently used in forensic laboratories. While this work focused on reducing the instrumental analysis portion of the fire debris analytical workflow, the implementation of less time-intensive sample preparation procedures is required to achieve substantial time savings. Thus, future investigations will focus on optimization of sampling procedures that utilize the rapid GC-MS probe holder as well as other approaches (e.g., solid-phase microextraction) to decrease sample preparation times.

Supplementary Material

Supp1

Highlights.

  • Rapid GC-MS screening technique applied to analysis of ignitable liquids.

  • Chromatographic analysis method optimized for volatile sample matrices.

  • Limits of detection for this method were 0.012 mg/mL to 0.018 mg/mL.

  • Major compounds in gasoline and diesel fuel identified in simulated burn samples.

Footnotes

Disclaimer

Certain commercial products are identified in order to adequately specify the procedure; this does not imply endorsement or recommendation by NIST, nor does it imply that such products are necessarily the best available for the purpose.

5. References

  • [1].Bell S, Forensic Chemistry, 2nd ed., Pearson Education, Glenview, IL, 2013. [Google Scholar]
  • [2].Stauffer E, Dolan JA, Newman R, Fire Debris Analysis, Academic Press, Burlington, MA, 2008. [Google Scholar]
  • [3].ASTM E1618-19 Standard Test Method for Ignitable Liquid Residues in Extracts from Fire Debris Samples by Gas Chromatography-Mass Spectrometry, ASTM International, West Conshohocken, PA, (2019). [Google Scholar]
  • [4].ASTM E1412-19 Standard Practice for Separation of Ignitable Liquid Residues from Fire Debris Samples by Passive Headspace Concentration with Activated Charcoal, ASTM International, West Conshohocken, PA: (2019). [Google Scholar]
  • [5].Ferreiro-González M, Ayuso J, Álvarez JA, Palma M, Barroso CG, Application of an HS–MS for the detection of ignitable liquids from fire debris, Talanta. 142 (2015) 150–156, 10.1016/j.talanta.2015.04.030. [DOI] [PubMed] [Google Scholar]
  • [6].Ferreiro-González M, Barbero GF, Ayuso J, Álvarez JA, Palma M, Barroso CG, Validation of an HS-MS method for direct determination and classification of ignitable liquids, Microchem. J. 132 (2017) 358–364, 10.1016/j.microc.2017.02.022. [DOI] [Google Scholar]
  • [7].Falatová B, Ferreiro-González M, Calle JLP, Álvarez JÁ, Palma M, Discrimination of Ignitable Liquid Residues in Burned Petroleum-Derived Substrates by Using HS-MS eNose and Chemometrics, Sensors. 21 (2021) 801, 10.3390/s21030801. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [8].Aliaño-González M, Ferreiro-González M, Barbero G, Palma M, Barroso C, Application of headspace gas chromatography-ion mobility spectrometry for the determination of ignitable liquids from fire debris, Separations. 5 (2018) 41, 10.3390/separations5030041. [DOI] [Google Scholar]
  • [9].Aliaño-González MJ, Ferreiro-González M, Barbero GF, Palma M, Novel method based on ion mobility spectrometry sum spectrum for the characterization of ignitable liquids in fire debris, Talanta. 199 (2019) 189–194, 10.1016/j.talanta.2019.02.063. [DOI] [PubMed] [Google Scholar]
  • [10].Barnett I, Bailey FC, Zhang M, Detection and classification of ignitable liquid residues in the presence of matrix interferences by using direct analysis in real time mass spectrometry, J. Forensic Sci. 64 (2019) 1486–1494, 10.1111/1556-4029.14029. [DOI] [PubMed] [Google Scholar]
  • [11].Sisco E, Forbes TP, Forensic applications of DART-MS: A review of recent literature, Forensic Chem. 22 (2021) 100294, 10.1016/j.forc.2020.100294. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [12].Liang J, Frazier J, Benefield V, Chong NS, Zhang M, Forensic fiber analysis by thermal desorption/pyrolysis-direct analysis in real time-mass spectrometry, Anal. Chem. 92 (2020) 1925–1933, 10.1021/acs.analchem.9b04167. [DOI] [PubMed] [Google Scholar]
  • [13].Fast analysis workflow with no sample preparation for forensic applications using Quickprobe GC-MS, Agilent Technologies, White Paper, (2019). [Google Scholar]
  • [14].Ignitable Liquids Reference Collection Database [internet] National Center for Forensic Science, University of Central Florida. [cited 02/23/22], Available from: http://ilrc.ucf.edu/. [Google Scholar]
  • [15].Capistran BA, McGuffin VL, Waddell Smith R, Application of a kinetic model to predict extracted ion profiles for the identification of evaporated ignitable liquids, Forensic Chem. 24 (2021) 100340, 10.1016/j.forc.2021.100340. [DOI] [Google Scholar]
  • [16].ASTM E2677 Limit of Detection Web Portal Data Entry Page [internet] National Institute of Standards and Technology. [cited 01/15/22], Available from: https://www-s.nist.gov/loda/index.html.
  • [17].Maximizing Efficiency of the Agilent QuickProbe GC/MS System, Agilent Technologies, White Paper, (2019). [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supp1

RESOURCES