. 2022 Aug 27;33(2):1229–1242. doi: 10.1007/s00330-022-09059-w

Table 2.

Artefacts induced by the generator per CIED model

n (%) PM 135 (78.9%) ICD 22 (12.9%) CRT-D 9 (5.3%) CRT-P 5 (2.9%) Total n = 171
n Median (IQR) n Median (IQR) n Median (IQR) n Median (IQR) p valuea p values of post hoc comparisons
Susceptibility signal-void artefact diameter measured from coronal localiser (bSSFP), cm 126 8.9 (7.9–9.8) 19 13.6 (12.6–14.5) 7 14.1 (13.0–14.8) 5 9.5 (9.0–9.7) < 0.001 < 0.001 (PM–ICD) < 0.001 (PM–CRT-D) 0.045 (CRT-P–ICD)
Banding artefact distance measured from coronal localiser (bSSFP), cm 126 10.4 (8.9–12.3) 19 20.6 (18.2–21.7) 7 18.8 (18.2–24.6) 5 11.8 (10.3–13.3) < 0.001 < 0.001 (PM–ICD) < 0.001 (PM–CRT-D)
Generator distance from the heart, measured from coronal localiser (bSSFP), cm 126 10.3 (8.6–12.0) 19 10.0 (8.5–11.9) 7 10.5 (7.9–11.3) 5 7.3 (6.5–10.7) 0.341

aKruskal-Wallis test p value

bSSFP balanced steady-state free precession

CIED cardiac implantable electronic device

CRT-D cardiac resynchronization therapy defibrillator

CRT-P cardiac resynchronization therapy pacemaker

ICD implantable cardioverter-defibrillator

LGE late-gadolinium enhancement

IQR interquartile range

PM pacemaker