Table 3.

Right ventricular pacing lead-induced artefact area on CMR per CIED model and sequence type

n (%) PM 135 (78.9%) ICD 22 (12.9%) CRT-D 9 (5.3%) CRT-P 5 (2.9%) Total n = 171
n Median (IQR) n Median (IQR) n Median (IQR) n Median (IQR) p value
bSSFP
  Cine SAX RV lead artefact area (cm2) 115 1.6 (1.2–2.1) 5 1.9 (1.1–2.3) 4 2.3 (2.1–3.5) 4 1.6 (1.1–1.8) 0.118
  LGE SAX RV lead artefact area (cm2) 110 1.6 (1.3–2.1) 2 1.2 5 1.9 (1.9–2.2) 1 1.1 0.160
  PSIR SAX RV lead artefact area (cm2) 113 1.6 (1.2–2.1) 4 1.2 (0.9–1.7) 6 1.9 (1.5–2.4) 3 1.1 0.283
SPGR
  Cine SAX RV lead artefact area (cm2) 21 1.8 (1.3–2.4) 18 1.4 (1.1–1.8) 4 1.4 (1.3–2.3) 3 2.0 0.083
  LGE SAX RV lead artefact area (cm2) 14 2.1 (1.3–2.7) 16 1.8 (1.4–2.1) 3 2.3 2 2.5 0.399
  PSIR SAX RV lead artefact area (cm2) 16 1.8 (1.3–2.5) 16 1.8 (1.2–2.3) 2 1.6 2 1.3 0.759

bSSFP balanced steady-state free precession

CIED cardiac implantable electronic device

CMR cardiac magnetic resonance imaging

CRT-D cardiac resynchronization therapy defibrillator

CRT-P cardiac resynchronization therapy pacemaker

ICD implantable cardioverter-defibrillator

LGE late-gadolinium enhancement

PM pacemaker

PSIR phase-sensitive inversion recovery

RV right ventricular

SAX short axis

SPGR spoiled gradient echo