Table 3.

The precision of measured concentrations of 232Th for QC and spiked urine materials (ng/L)

Sample N (number of runs) Average concentration (ng/L) SD RSD (%) Target value c (ng/L) Bias (%)

Low QCa 60 47.2 2.68 5.7 50 −5.6
High QCa 60 741 31.0 4.2 750 −1.2
Pool-1b 16 39.1 1.6 4.1 40 −2.3
Pool-2 b 16 194 4.1 2.1 200 −3.0
a

Internal quality control material made at CDC by spiking certified reference material in pooled urine collected anonymously

b

Urine materials made at CDC by spiking certified reference material into pooled urine collected anonymously

c

Target values of spiked QCs and urine pools using NIST traceable certified reference material from High Purity Standards (Charleston, SC) and Inorganic Ventures (Christiansburg, VA)