Table 2.
Pearson’s correlation and within-person coefficient of variation (CVw) for CGM parameters from the Abbott Libre Pro and Dexcom G4 Platinum sensors comparing two-week wear periods, assessed at baseline and the 3-month follow-up visit, N=153
| Abbott Libre Pro | Dexcom G4 Platinum | |||
|---|---|---|---|---|
| Pearson’s Correlation (r) |
CVw (95% CI)* | Pearson’s Correlation (r) |
CVw (95% CI)* | |
| Mean glucose, mg/dL | 0.69 | 17.4% (14.4, 20.4) | 0.73 | 14.2% (12.0, 16.5) |
| Percent time-in-range glucose 70-180 mg/dL | 0.71 | 20.1% (16.9, 23.2) | 0.73 | 18.6% (15.6, 21.6) |
| Percent of time above glucose >180 mg/dL | 0.72 | 53.0% (44.4, 61.7) | 0.71 | 52.5% (41.5, 63.5) |
| Percent of time above glucose >250 mg/dL | 0.70 | 122.5% (85.7, 159.3) | 0.72 | 118.1% (90.3, 146.0) |
| Percent of time below glucose <70 mg/dL | 0.38 | 157.8% (121.2, 194.3) | 0.43 | 160.6% (93.9, 227.3) |
| Percent of time below glucose <54 mg/dL | 0.26 | 263.5% (187.7, 339.2) | 0.24 | 298.8% (166.3, 431.3) |
The 95% CIs were bootstrapped with 200 replications