Table 4.

GC/MS operating conditions for PAHs analysis.

Parameters Conditions
Gas Chromatography
Column Agilent Technology, DB-5 ms (0–325 °C; 30 m*250 μm*0.25 μm)
Carrier Gas Helium
Injection Volume
Flow Rate
1 μL, Splitless
1 mL/min, Constant Flow

Temperature Program
Initial Temperature 70 °C, hold 4 min
Final Temperature 300 °C (20 °C/min to 120 °C, and then 10 °C/min to 300 °C), hold 10 min.
Total Run Time 34.5 min

Mass Spectrometer
Transfer Line Temperature 290 °C
Source Temperature 230 °C
Electron Energy 70 volts
Ionization Mode Electron ionization (EI)
Mass Range 40–500 amu, SIM Mode, Time Segments