Differentials of diagnostic sensitivity for the participating Cytologists 1, 2, and 3, diagnoses committed through BestCyte Cell Sorter Imaging for the incremental positive thresholds: ASCUS+, LSIL+, and ASCH+, as defined by Truth Reference Diagnoses, including differences in sensitivity values, 95% Confidence Interval lower and upper limits, and equivalency grading relative to manual microscopy predicate diagnoses.

Diagnostic sensitivity
95% Confidence interval
Equivalency
Diagnosis threshold Cytologist Manual microscopy Cytologist Difference Lower limit Upper limit Grading
ASCUS+ 1 0.9136 0.9588 –0.0453 –0.0852 –0.0053 S. Superior
2 0.9136 0.9547 –0.0412 –0.0709 –0.0114 S. Superior
3 0.9136 0.9465 –0.0329 –0.0784 0.0125 Superior
LSIL+ 1 0.8667 0.9697 –0.1030 –0.1524 0.0537 S. Superior
2 0.8667 0.9576 –0.0909 –0.1487 0.0332 S. Superior
3 0.8667 0.9515 –0.0848 –0.1464 0.0233 S. Superior
ASCH+ 1 0.7436 0.9487 –0.2051 –0.3015 0.1087 S. Superior
2 0.7436 0.9359 –0.1923 –0.3105 0.0741 S. Superior
3 0.7436 0.9872 –0.2436 –0.3453 0.1419 S. Superior