Skip to main content
. 2023 Jan 18;8(1):45–50. doi: 10.5194/jbji-8-45-2023

Table 3.

Test accuracy of the BIOFIRE JI Panel for infection according to criteria and cultures (Methods section) as reference standard in patients with a clinical suspicion of native septic arthritis, early, or late acute PJIs.

All samples ( n=45 ) 95 % CI Native septic arthritis 95 % CI Early acute PJI 95 % CI Late acute PJI 95 % CI
      ( n=12 )   ( n=14 )   ( n=19 )  
% infection diagnosis 6/12 (50 %)   6/12 (50 %)   10/14 (71 %)   15/19 (80 %)  
% BIOFIRE correct 73.3 %   91.7 %   50.0 %   78.9 %  
Sensitivity 61.3 % 44–76 83.3 % 44–97 30.0 % 11–60 73.3 % 48–89
Specificity 100 % 78–100 100 % 61–100 100 % 51–100 100 % 51–100